FDA Adverse Event Other Summary report: N

DRAEGER APOLLO

MDR report key: 1151945 · Received September 5, 2008

Report

Report Number
MW5008215
Event Type
Other
Date Received
September 5, 2008
Date of Event
August 20, 2008
Report Date
September 5, 2008
Manufacturer
NORTH AMERICAN DRAEGER
Product Code
BSZ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

DRAEGER ANESTHESIA MACHINE, MODEL APOLLO. GAS SAMPLE LINE WATER TRAP PLUGGED CAUSING ETCO2 TO GO TO UNACCEPTABLE LEVEL. REPEATED INTERMITTENT ALARMS "SAMPLE LINE PLUGGED" WERE NOT RESOLVED PRIOR TO PROBLEM, ALTHOUGH LABELING OF ALARM WAS MISLEADING. MACHINE SWAPPED IN USE, NO NOTED PT. ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRAEGER APOLLO ANESTHESIA MACHINE BSZ NORTH AMERICAN DRAEGER APOLLO

Patients

Seq Age Sex Outcome Treatment
1 * Other