FDA Adverse Event
Other
Summary report: N
DRAEGER APOLLO
MDR report key: 1151945
·
Received September 5, 2008
Report
- Report Number
- MW5008215
- Event Type
- Other
- Date Received
- September 5, 2008
- Date of Event
- August 20, 2008
- Report Date
- September 5, 2008
- Manufacturer
- NORTH AMERICAN DRAEGER
- Product Code
- BSZ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
DRAEGER ANESTHESIA MACHINE, MODEL APOLLO. GAS SAMPLE LINE WATER TRAP PLUGGED CAUSING ETCO2 TO GO TO UNACCEPTABLE LEVEL. REPEATED INTERMITTENT ALARMS "SAMPLE LINE PLUGGED" WERE NOT RESOLVED PRIOR TO PROBLEM, ALTHOUGH LABELING OF ALARM WAS MISLEADING. MACHINE SWAPPED IN USE, NO NOTED PT. ISSUES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DRAEGER APOLLO | ANESTHESIA MACHINE | BSZ | NORTH AMERICAN DRAEGER | APOLLO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |