FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT II SYRINGE

MDR report key: 11518130 · Received March 18, 2021

Report

Report Number
3002682307-2021-00097
Event Type
Malfunction
Date Received
March 18, 2021
Date of Event
February 15, 2021
Report Date
May 3, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2011154. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE ORIGINALLY FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS ISSUE, THE AFFECTED SAMPLE WAS RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE SAMPLE, LEAKAGE WAS OBSERVED THROUGH THE PLUNGER ROD. THROUGH MAGNIFIED INSPECTION, DAMAGE WAS OBSERVED TO THE PLUNGER LIP COMPONENT. THIS TYPE OF DAMAGE CAN BE PRODUCED DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR WITHIN THE PLUNGER ASSEMBLY MACHINE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE LEAKED PAST THE STOPPER DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH TO ENGLISH: "LACK OF TIGHTNESS AT THE PISTON LEVEL, THE LIQUID LEAKS ALONG. CLINICAL CONSEQUENCE: NONE THIS INCIDENT TOOK PLACE IN HEMATOLOGY ON (B)(6) 2021."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD DISCARDIT" II SYRINGE LEAKED PAST THE STOPPER DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "LACK OF TIGHTNESS AT THE PISTON LEVEL, THE LIQUID LEAKS ALONG. CLINICAL CONSEQUENCE: NONE. THIS INCIDENT TOOK PLACE IN HEMATOLOGY ON (B)(6) 2021."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411124 BD DISCARDIT II SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2011154

Patients

Seq Age Sex Outcome Treatment
1