INFUSOMAT
Report
- Report Number
- 9610825-2021-00100
- Event Type
- Malfunction
- Date Received
- March 18, 2021
- Report Date
- August 4, 2021
- Manufacturer
- B.BRAUN MELSUNGEN AG
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MELSUNGEN AG INTERNAL REPORT (B)(4). THE DEVICE WAS INVESTIGATED IN OUR LABORATORY IN SAO GONCALO, BRAZIL: ACCORDING TO THE EQUIPMENT'S HISTORY, WE FOUND ONLY A SINGLE 330ML SCHEDULE, ON (B)(6) 2020, AT 13:50:40, COMPATIBLE WITH THE CUSTOMER'S INFORMATION. HOWEVER, THE TOTAL PROGRAMMED INFUSION TIME WAS FROM 5:30 PM TO 1:50:51 PM, RESULTING IN A PROGRAMMED FLOW RATE OF 18.86 ML/H. THE INFUSION TOOK PLACE EXACTLY AS SCHEDULED UNTIL (B)(6) 2020, AT 06:10:28, BY THE USER. THE TOTAL AMOUNT OF DRUG INFUSED WAS 301.39ML. RESULTING IN A TOTAL INFUSION TIME OF 16:18:46. WE VERIFIED IN THE PERIOD ONLY TWO ALARMS THAT WERE INHIBITED IN MINUTES THE EQUIPMENT WORKED AS EXPECTED, THE DEVIATION INFORMATION PROVIDED BY THE CUSTOMER WAS NOT CONFIRMED IN THE OPERATING HISTORY. UNCONFIRMED TECHNICAL COMPLAINT.
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN (B)(4) AG INTERNAL REPORT (B)(4). THE DEVICE HAS BEEN REQUESTED TO BE SENT FOR AN EXAMINATION AND ROOT CAUSE ANALYSIS IN OUR SERVICE LABORATORY IN (B)(4). A FOLLOW-UP REPORT WILL BE PROVIDED AS SOON AS RESULTS ARE AVAILABLE. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN (B)(4): "OVERINFUSION". ACCORDING TO CUSTOMER: "AV PATIENT'S NPP ENDED AT 4:30, BUT STARTED TO ALARM AT 4 O'CLOCK AND TOOK THE WHOLE SCHOLARSHIP INSTEAD OF THE 330 SCHEDULE, AND IT WAS ONLY GOING TO END AT 6:40 AM ON THE SCHEDULE." "
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 416215 | INFUSOMAT | PUMP, INFUSION | FRN | B.BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |