FDA Adverse Event
Injury
Summary report: N
CONMED ELECTROSURGICAL DISPERSIVE ELECTRODE
MDR report key: 115111
·
Received August 25, 1997
Report
- Report Number
- 1317214-1997-00103
- Event Type
- Injury
- Date Received
- August 25, 1997
- Date of Event
- July 23, 1997
- Report Date
- July 28, 1997
- Manufacturer
- CONMED CORPORATION
- Product Code
- GEI
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT A MALE PT REC'D 1 SQ. CM. OF BURNING, 2ND DEGREE, DURING AN OPERATION OF A TIBIOFIBULAR ON THE RIGHT LEG. THE SPOT TURNED BLACK AND CARBONIZED. A NURSE PUT THIS GROUND PAD ON THE LEFT THIGH WITHOUT SHAVING EXCESSIVELY HAIRY. AT THE POWER OF 40W, THIS PT COMPLAINED ABOUT NUMBNESS. WHEN THE ESU POWER WAS RAISED TO 50W, HE COMPLAINED OF STRONG PAD AT THE AREA OF THE GROUND PAD. THE NURSE FOUND THE BURNING, THE GROUND PAD WAS PLACED PROPERLY AND NOTHING WRONG WAS FOUND ON THE GROUND PAD. THIS MACHINE HAS A FUNCTION/CAPABILITY OF ALARM BUT DID NOT ALARM DURING THE OPERATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONMED ELECTROSURGICAL DISPERSIVE ELECTRODE | DISPERSIVE ELECTORDE | GEI | CONMED CORPORATION | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Required Intervention | ESU |