FDA Adverse Event Injury Summary report: N

CONMED ELECTROSURGICAL DISPERSIVE ELECTRODE

MDR report key: 115111 · Received August 25, 1997

Report

Report Number
1317214-1997-00103
Event Type
Injury
Date Received
August 25, 1997
Date of Event
July 23, 1997
Report Date
July 28, 1997
Manufacturer
CONMED CORPORATION
Product Code
GEI
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT A MALE PT REC'D 1 SQ. CM. OF BURNING, 2ND DEGREE, DURING AN OPERATION OF A TIBIOFIBULAR ON THE RIGHT LEG. THE SPOT TURNED BLACK AND CARBONIZED. A NURSE PUT THIS GROUND PAD ON THE LEFT THIGH WITHOUT SHAVING EXCESSIVELY HAIRY. AT THE POWER OF 40W, THIS PT COMPLAINED ABOUT NUMBNESS. WHEN THE ESU POWER WAS RAISED TO 50W, HE COMPLAINED OF STRONG PAD AT THE AREA OF THE GROUND PAD. THE NURSE FOUND THE BURNING, THE GROUND PAD WAS PLACED PROPERLY AND NOTHING WRONG WAS FOUND ON THE GROUND PAD. THIS MACHINE HAS A FUNCTION/CAPABILITY OF ALARM BUT DID NOT ALARM DURING THE OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONMED ELECTROSURGICAL DISPERSIVE ELECTRODE DISPERSIVE ELECTORDE GEI CONMED CORPORATION NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Required Intervention ESU