SYRINGE 0.3ML 31GA 8MM HALFUNIT 10BG 500
Report
- Report Number
- 1920898-2021-00310
- Event Type
- Malfunction
- Date Received
- March 17, 2021
- Date of Event
- February 19, 2021
- Report Date
- March 19, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00382908440035
- PMA / PMN Number
- K024112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED (34) LOOSE 3/10CC, 8MM SYRINGES. CUSTOMER STATES THAT THE SCALE PRINT SMUDGED. ALL RETURNED SYRINGES WERE EXAMINED AND ALL EXHIBITED SMUDGED INK WHERE SCALE MARKINGS WERE MISSING BETWEEN THE 0-10 UNIT MARKINGS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0203537. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS EXCEPT AS NOTED BELOW. THERE WAS ONE (1) NOTIFICATION NOTED FOR DAMAGED TIPS CAUSING MISSING PRINT. ROOT CAUSE: TIMING WAS OFF CAUSING DAMAGED BARREL TIPS AND PRINT ISSUES. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT SYRINGE 0.3ML 31GA 8MM HALFUNIT 10BG 500 HAD SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 328440 BATCH NO: 0203537. IT WAS REPORTED SCALE PRINT SMUDGED. VERBATIM: EMAIL RECEIVED, SCALE PRINT SMUDGED.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT SYRINGE 0.3ML 31GA 8MM HALFUNIT 10BG 500 HAD SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 328440 BATCH NO: 0203537. IT WAS REPORTED SCALE PRINT SMUDGED. VERBATIM: EMAIL RECEIVED, SCALE PRINT SMUDGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403971 | SYRINGE 0.3ML 31GA 8MM HALFUNIT 10BG 500 | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 328440 | 0203537 | 00382908440035 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |