FDA Adverse Event Malfunction Summary report: N

HS HAND CONTROLLED CURVED SHEAR

MDR report key: 1150973 · Received August 28, 2008

Report

Report Number
3005075853-2008-01390
Event Type
Malfunction
Date Received
August 28, 2008
Date of Event
July 23, 2008
Report Date
July 24, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GEI
PMA / PMN Number
K002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DURING FUNCTIONAL TESTING, AN ERROR CODE WAS DISPLAYED AND THE BLADE TIP FURTHER CRACKED. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. WHEN A BLADE HAS BEEN COMPROMISED SUCH AS SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. A POSSIBLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE. EACH DEVICE IS VISUALLY INSPECTED, AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ROUX-EN-Y PROCEDURE, THE DEVICE GAVE AN ERROR 5. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HS HAND CONTROLLED CURVED SHEAR GEI ETHICON ENDO-SURGERY, LLC NA UNK

Patients

Seq Age Sex Outcome Treatment
1 HANDPIECE| GENERATOR