FDA Adverse Event Injury Summary report: N

NA

MDR report key: 1150519 · Received September 4, 2008

Report

Report Number
1056128-2008-00059
Event Type
Injury
Date Received
September 4, 2008
Date of Event
August 10, 2008
Report Date
September 4, 2008
Manufacturer
ASCENT HEALTHCARE SOLUTIONS
Product Code
NLQ
PMA / PMN Number
K043315
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE RETURNED DEVICE SHOWED TISSUE BUILD-UP AS WELL AS CHARRING ON THE BLADE OF THE DEVICE. ASCENT'S INSTRUCTIONS FOR USE STATES: "BLOOD AND TISSUE BUILDUP BETWEEN THE BLADE AND SHAFT MAY RESULT IN ABNORMALLY HIGH TEMPERATURES AT THE DISTAL END OF THE SHAFT. TO PREVENT BURN INJURY, REMOVE ANY VISIBLE TISSUE BUILDUP AT THE DISTAL END OF THE SHAFT. DURING PROLONGED ACTIVATION, THE BLADE, THE CLAMP ARM AND THE DISTAL 7CM OF THE SHAFT MAY BECOME HOT. AVOID UNINTENDED CONTACT WITH TISSUE, DRAPES, SURGICAL GOWNS, OR OTHER UNINTENDED SITES AT ALL TIMES." A REVIEW OF THE LOT CONTROL SHEET FOR THE REPORTED DEVICE SERIAL NUMBER, INDICATED THAT THE INSTRUMENT PASSED ALL APPLICABLE TESTS AND INSPECTIONS PRIOR TO RELEASE.

Description of Event or Problem · 1

DURING THE PROCEDURE, THE HARMONIC SCALPEL DEVICE WAS BURNING TISSUE THAT CAME INTO CONTACT WITH THE SHAFT OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NA NLQ ASCENT HEALTHCARE SOLUTIONS ACE36P 175363

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention