FDA Adverse Event Injury Summary report: N

TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET

MDR report key: 1150104 · Received August 29, 2008

Report

Report Number
3005099803-2008-04207
Event Type
Injury
Date Received
August 29, 2008
Date of Event
May 13, 2008
Report Date
August 1, 2008
Manufacturer
BOSTON SCIENTIFIC
Product Code
LQC
PMA / PMN Number
k040447
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVAL, THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PROD FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFO FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOSCOPY PROCEDURE (ERCP), A TIP BREAKAGE OCCURRED. THE PROCEDURE WAS BEING PERFORMED FOR REMOVAL OF GALLSTONES FROM AN UNSPECIFIED LOCATION. TWO LARGE STONES WERE REMOVED. UPON ADVANCING THE RX LITHOTRIPTER COMPATIBLE BASKET IN AN ATTEMPT TO RETRIEVE A THIRD STONE, THE METAL TIP BROKE AND REMAINED INSIDE THE PT. THE PHYSICIAN PLACED AN UNSPECIFIED PLASTIC BILIARY STENT FOR THE DUCT TO REMAIN PATENT AND DRAINING, DESIRING THAT THE METAL TIP WILL DRAIN OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRAPEZOID RX LITHOTRIPTER COMPATIBLE BASKET LQC LQC BOSTON SCIENTIFIC M00510880 11289956

Patients

Seq Age Sex Outcome Treatment
1 * Required Intervention