PUMP MMT-512NAB PRDGM INSULIN BL EN

MDR report key: 1150034 · Received August 27, 2008

Report

Report Number
2032227-2008-01468
Event Type
Malfunction
Date Received
August 27, 2008
Date of Event
August 13, 2008
Report Date
August 13, 2008
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
K030531
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS, AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER COMPLAINED OF HIGH BLOOD GLUCOSE LEVELS. THE REPORTED BLOOD GLUCOSE LEVEL WAS 375 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE PRIME TEST. THE PRIME TEST WAS PERFORMED A SECOND TIME WITH A NEW RESERVOIR AND INFUSION SET, AND THE INSULIN PUMP PASSED. HOWEVER, THE INSULIN PUMP FAILED THE HIGH PRESSURE TEST TWICE. NOTHING FURTHER WAS REPORTED.