FDA Adverse Event Injury Summary report: N

BD E-Z SCRUB

MDR report key: 11499207 · Received March 16, 2021

Report

Report Number
1423507-2021-00014
Event Type
Injury
Date Received
March 16, 2021
Date of Event
February 4, 2021
Report Date
August 3, 2021
Manufacturer
CAREFUSION, INC
Product Code
GEC
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

FOLLOW UP EMDR REVIEW OF THE MANUFACTURING BATCH RECORD AND LABORATORY RESULTS DID NOT DETERMINE ROOT CAUSE FOR THIS ISSUE. REVIEW OF PREVIOUS COMPLAINTS DID NOT IDENTIFY PREVIOUS COMPLAINTS FOR THIS LOT. REVIEW OF THE FINISHED GOOD BATCH RECORD, PART NUMBER 371163, LOT 0133746, WAS PERFORMED FOR ANY DISCREPANCIES OR DEVIATIONS, INCLUDING EXCESSIVE SOAP WEIGHT, THAT WOULD CONTRIBUTE TO THIS ISSUE. NO DEVIATIONS OR DISCREPANCIES WERE IDENTIFIED DURING THE REVIEW OF THE BATCH RECORD. REVIEW OF VENDOR¿S COA AND BD SANDY COA DID NOT IDENTIFY ANY OUT OF SPECIFICATIONS OR DISCREPANCIES THAT WOULD CONTRIBUTE TO THIS ISSUE.

Description of Event or Problem · 0

THE CENTRUM DISTRIBUTOR REPORTS THAT SURGEON LSMF FROM THE COLON AND RECTUM AREA OF THE CENTRAL MILITARY HOSPITAL REPORTED THAT SHE HAS PRESENTED ERYTHEMA ON HER HANDS AFTER USING THE SURGICAL BRUSH CATALOG 371163 (PCMX 3%) FOR WHICH SHE REQUESTS THE CHANGE OF CATALOG TO 371073 (CHG 4%). IN ADDITION, SHE STATES THAT THE EPIDEMIOLOGY AREA OF THE HOSPITAL CARRIED OUT A STUDY WHERE THE PRESENCE OF MICROBIOTA WAS EVEN DETECTED IN THE AREA ALREADY WASHED WITH THE BRUSH. THIS CLAIM WAS OPENED TO INVESTIGATE THE PRESENCE OF MICROBIOTA. COMPLAINT (B)(4) WAS OPENED TO INVESTIGATE ERYTHEMA.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CENTRUM DISTRIBUTOR REPORTS THAT SURGEON LSMF FROM THE COLON AND RECTUM AREA OF THE CENTRAL MILITARY HOSPITAL REPORTED THAT SHE HAS PRESENTED ERYTHEMA ON HER HANDS AFTER USING THE SURGICAL BRUSH CATALOG 371163 (PCMX 3%) FOR WHICH SHE REQUESTS THE CHANGE OF CATALOG TO 371073 (CHG 4%). IN ADDITION, SHE STATES THAT THE EPIDEMIOLOGY AREA OF THE HOSPITAL CARRIED OUT A STUDY WHERE THE PRESENCE OF MICROBIOTA WAS EVEN DETECTED IN THE AREA ALREADY WASHED WITH THE BRUSH. THIS CLAIM WAS OPENED TO INVESTIGATE THE PRESENCE OF MICROBIOTA. COMPLAINT (B)(4) WAS OPENED TO INVESTIGATE ERYTHEMA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
392599 BD E-Z SCRUB IMPREGNATED PREOPERATIVE SCRUB BRUSHES (3% CHLOROXYLENOL (PCMX)) GEC CAREFUSION, INC 0133746

Patients

Seq Age Sex Outcome Treatment
1 Other