FDA Adverse Event Malfunction Summary report: N

VACUETTE MULTIPLE USE DRAWING NEEDLE 21G X 1 1/4" GREEN, STERILE

MDR report key: 11495776 · Received March 16, 2021

Report

Report Number
8020040-2021-00013
Event Type
Malfunction
Date Received
March 16, 2021
Report Date
April 7, 2021
Manufacturer
GREINER BIO-ONE GMBH
Product Code
FMI
PMA / PMN Number
K973620
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

RECEIVED 158PC 450062/19H08C FOR EVALUATION. RECEIVED CUSTOMER PICTURE. WE HAVE NO FURTHER INVENTORY OF THE MATERIAL/BATCH. WE HAVE NO FURTHER COMPLAINTS ON THE MATERIAL/BATCH. WE FORWARDED THE COMPLAINT, CUSTOMER SAMPLES AND PICTURE TO OUR SUPPLIER FOR THEIR INVESTIGATION AND COMMENTS. A REVIEW OF THE MANUFACTURING AND INSPECTION RECORDS OF THE CONCERNED MATERIAL/BATCH SHOWED NO ABNORMALITY WHICH COULD BE RELATED TO THE REPORTED ISSUE. RETAIN AND CUSTOMER SAMPLES WERE VISUALLY INSPECTED; NO BENT NOR BROKEN FRONT NEEDLE WAS FOUND. NEXT, THE RIGIDITY OF THE NEEDLE TUBES WAS TESTED; ALL WERE WITHIN THE RANGE OF STANDARDS. THE COMPLAINT COULD NOT BE DUPLICATED. CORRECTED DATA: H3: DEVICE EVALUATED BY MANUFACTURER; H6: INVESTIGATION METHOD, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSION.

Additional Manufacturer Narrative · 1

COMPLAINT (B)(4). A DATE OF THE EVENT COULD NOT BE OBTAINED FROM THE CUSTOMER. THE SAMPLES WERE JUST RECENTLY OBTAINED AND FORWARDED TO THE SUPPLIER OF THE PRODUCT FOR AN EVALUATION. AS SOON AS THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

CUSTOMER STATES NEEDLES BROKE OFF; ONE PRIOR TO DRAW AND THE OTHER IN THE MIDDLE OF ENGAGING. CLARIFICATION HAS BEEN REQUESTED ON "IN THE MIDDLE OF ENGAGING". PRODUCT SPECIALIST HAS BEEN MADE AWARE OF THE INCIDENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396808 VACUETTE MULTIPLE USE DRAWING NEEDLE 21G X 1 1/4" GREEN, STERILE CANNULA FMI GREINER BIO-ONE GMBH 450062 19H08C

Patients

Seq Age Sex Outcome Treatment
1