VACUETTE MULTIPLE USE DRAWING NEEDLE 21G X 1 1/4" GREEN, STERILE
Report
- Report Number
- 8020040-2021-00013
- Event Type
- Malfunction
- Date Received
- March 16, 2021
- Report Date
- April 7, 2021
- Manufacturer
- GREINER BIO-ONE GMBH
- Product Code
- FMI
- PMA / PMN Number
- K973620
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
RECEIVED 158PC 450062/19H08C FOR EVALUATION. RECEIVED CUSTOMER PICTURE. WE HAVE NO FURTHER INVENTORY OF THE MATERIAL/BATCH. WE HAVE NO FURTHER COMPLAINTS ON THE MATERIAL/BATCH. WE FORWARDED THE COMPLAINT, CUSTOMER SAMPLES AND PICTURE TO OUR SUPPLIER FOR THEIR INVESTIGATION AND COMMENTS. A REVIEW OF THE MANUFACTURING AND INSPECTION RECORDS OF THE CONCERNED MATERIAL/BATCH SHOWED NO ABNORMALITY WHICH COULD BE RELATED TO THE REPORTED ISSUE. RETAIN AND CUSTOMER SAMPLES WERE VISUALLY INSPECTED; NO BENT NOR BROKEN FRONT NEEDLE WAS FOUND. NEXT, THE RIGIDITY OF THE NEEDLE TUBES WAS TESTED; ALL WERE WITHIN THE RANGE OF STANDARDS. THE COMPLAINT COULD NOT BE DUPLICATED. CORRECTED DATA: H3: DEVICE EVALUATED BY MANUFACTURER; H6: INVESTIGATION METHOD, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSION.
COMPLAINT (B)(4). A DATE OF THE EVENT COULD NOT BE OBTAINED FROM THE CUSTOMER. THE SAMPLES WERE JUST RECENTLY OBTAINED AND FORWARDED TO THE SUPPLIER OF THE PRODUCT FOR AN EVALUATION. AS SOON AS THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
CUSTOMER STATES NEEDLES BROKE OFF; ONE PRIOR TO DRAW AND THE OTHER IN THE MIDDLE OF ENGAGING. CLARIFICATION HAS BEEN REQUESTED ON "IN THE MIDDLE OF ENGAGING". PRODUCT SPECIALIST HAS BEEN MADE AWARE OF THE INCIDENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 396808 | VACUETTE MULTIPLE USE DRAWING NEEDLE 21G X 1 1/4" GREEN, STERILE | CANNULA | FMI | GREINER BIO-ONE GMBH | 450062 | 19H08C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |