FDA Adverse Event Malfunction Summary report: N

RIGIDFIX MALLET *EA

MDR report key: 11486035 · Received March 15, 2021

Report

Report Number
1221934-2021-00845
Event Type
Malfunction
Date Received
March 15, 2021
Report Date
March 1, 2021
Manufacturer
DEPUY MITEK LLC US
Product Code
LXH
UDI-DI
10886705002993
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H10 ADDITIONAL NARRATIVE: D9, H3, H6: THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. INVESTIGATION SUMMARY BACKGROUND: THE MALLET¿S HANDLE HAS COME OFF IN SURGEON¿S HANDS DURING THE PROCEDURE. INVESTIGATION SUMMARY: THE RIGIDFIX MALLET *EA (PART #: 213705, LOT #: UNK) WAS RECEIVED AND EVALUATED AT US CQ. UPON VISUAL INSPECTION, THE MALLET HANDLE BROKE OFF THE HEAD. THERE WERE SCRATCHES OBSERVED ON THE BOTH THE FACES OF THE MALLET HEAD. HENCE, THE REPORTED CONDITION CAN BE CONFIRMED. THE POSSIBLE ROOT CAUSE FOR THE REPORTED CONDITION CAN BE THAT THE DEVICE ENCOUNTERED UNINTENDED FORCES DURING ITS USE. DURING THE INVESTIGATION, NO PRODUCT DESIGN OR MANUFACTURING ISSUES WERE OBSERVED THAT MAY HAVE CONTRIBUTED TO THE COMPLAINT CONDITION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT TO WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: B5: SUBSEQUENT FOLLOW-UP WITH THE CUSTOMER, ADDITIONAL INFORMATION WAS RECEIVED ABOUT THE EVENT; THEREFORE, THE EVENT DESCRIPTION HAS BEEN UPDATED TO REFLECT THIS ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED BY THE SALES REP IN UNITED KINGDOM THAT DURING AN ARTHROPLASTY PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE HANDLE ON THE RIGIDFIX MALLET DEVICE CAME OFF THE SURGEON'S HAND. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UDI: (B)(4). THE EXPIRATION DATE IS CURRENTLY UNAVAILABLE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE HANDLE ON THE RIGIDFIX MALLET DEVICE CAME OFF THE SURGEON'S HAND. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
384868 RIGIDFIX MALLET *EA SURGICAL MALLET LXH DEPUY MITEK LLC US 213705 10886705002993

Patients

Seq Age Sex Outcome Treatment
1