FDA Adverse Event Injury Summary report: N

REVERSE SHOULDER SYSTEM GLENOID BASEPLATE 27X15

MDR report key: 11473640 · Received March 12, 2021

Report

Report Number
3005180920-2021-00191
Event Type
Injury
Date Received
March 12, 2021
Date of Event
February 5, 2021
Report Date
June 2, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
PHX
UDI-DI
07630040706414
PMA / PMN Number
K170452
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTION PERFORMED BY R&D MANAGER THE BASEPLATE APPEARS TO GENERALLY HAVE HA RESIDUALS ON THE COATED SURFACE, WHILE SOME SPOTS PRESENT BONE RESIDUALS, FOLLOWING EARLY BONY RESPONSE. THE ARTICULAR SURFACE OF THE GLENOSPHERE IS PARTIALLY SCRATCHED DUE TO FRICTION WITH THE REVERSE METAPHYSIS FOLLOWING BASEPLATE MOBILIZATION AND SUBSEQUENT LUXATION. THE GLENOSPHERE SCREW AND THE GLENOID LOCKING SCREWS DO NOT SHOW ANY SIGNS OF DAMAGE. NO ACTION IS SUGGESTED.

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDACTA CLINICAL DIRECTOR: EARLY MOBILISATION OF THE GLENOID BASEPLATE IN PRIMARY RSA, FEW WEEKS AFTER INDEX SURGERY. THE ORIGINAL BASEPLATE HAD SHORT PEG, THIS CHOICE WAS MOST PROBABLY DONE TO PRESERVE BONE WITH A COMPROMISE ON HOLD, WHICH IS COMMON PRACTICE. NO REASON TO SUSPECT A FAULTY DEVICE.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 22 FEBRUARY 2021: LOT 2001981: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 29-APR-2020. EXPIRATION DATE: 2025-04-15. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

REVISION SURGERY 4 MONTHS AFTER PRIMARY FOR BASE PLATE LOOSENING. THE SURGEON REVISED SUCCESSFULLY THE PATIENT SHOULDER REPLACING THE GLENOIDS IMPLANTS AND THE HUMERAL REVERSE HC LINER. THE GLENOSPHERE WAS FOUND COUPLED TO THE BASEPLATE DURING THE REVISION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
368394 REVERSE SHOULDER SYSTEM GLENOID BASEPLATE 27X15 GLENOID BASEPLATE PHX MEDACTA INTERNATIONAL SA 04.01.0154 2001981 07630040706414

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention