FDA Adverse Event Malfunction Summary report: N

IDNOW COVID-19

MDR report key: 11467218 · Received March 11, 2021

Report

Report Number
1221359-2021-00733
Event Type
Malfunction
Date Received
March 11, 2021
Report Date
April 20, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION/CORRECTED DATA: B6. H10: THE CUSTOMER SENT A CUMULATIVE REPORT OF THIRTY (30) FALSE NEGATIVES OVER NINE (9) LOT NUMBERS. THIS MFR. REPORT ADDRESSES LOT FIVE (5) OF NINE (9). THE CUSTOMER REPORTED FIVE (5) FALSE NEGATIVE TEST RESULTS ON NASAL SWABS WITH THE IDNOW COVID-19 ASSAY. CONFIRMATION TESTING USING ONE (1) NASAL SWAB IN VIRAL TRANSPORT MEDIA WITH CORELAB PLATFORM AND FOUR (4) NASOPHARYNGEAL SWABS IN VIRAL TRANSPORT MEDIA WITH CORELAB PLATFORM GENERATED POSITIVE RESULTS (CT VALUES NOT PROVIDED). ADDITIONAL INFORMATION: D3, D4 (UDI), G1 H10: ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLE.

Additional Manufacturer Narrative · 1

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M133949 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000/ LOT: M133949, TEST BASE PART NUMBER 190-430 / LOT:M133949. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M133949 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. MFR REPORTS ASSOCIATED WITH THIS CASE: 1221359-2021-00642, 1221359-2021-00650, 1221359-2021-00651, 1221359-2021-00732, 1221359-2021-00654, 1221359-2021-00655, 1221359-2021-00656, 1221359-2021-00657.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FIVE FALSE NEGATIVE TEST RESULTS USING ID NOW COVID-19 ASSAY - ONE WITH NASAL SWABS IN (B)(6) LAB, FOUR WITH NASOPHARYNGEAL SWABS IN TRANSPORT (B)(6) LAB - PERFORMED ON TEST DATE (NOT PROVIDED). SAMPLES WERE COLLECTED (B)(6) 2021. CONFIRMATION TESTING USING NASOPHARYNGEAL SWABS WITH PCR ABBOTT M2000 GENERATED POSITIVE RESULTS (CT VALUES NOT PROVIDED). NO PATIENT INFORMATION, HISTORY, OR IMPACT IS AVAILABLE. THE ID NOW COVID-19 PRODUCT INSERT INDICATES THAT NEGATIVE RESULTS SHOULD BE TREATED AS PRESUMPTIVE AND TESTED WITH AN ALTERNATIVE FDA AUTHORIZED MOLECULAR ASSAY, IF NECESSARY FOR CLINICAL MANAGEMENT, INCLUDING INFECTION CONTROL. THE PI STATES NEGATIVE RESULTS DO NOT PRECLUDE SARS-C0V-2 INFECTION AND SHOULD NOT BE USED AS THE SOLE BASIS FOR PATIENT MANAGEMENT DECISIONS. NEGATIVE RESULTS MUST BE COMBINED WITH CLINICAL OBSERVATIONS, PATIENT HISTORY, AND EPIDEMIOLOGICAL INFORMATION. ADDITIONALLY, THE ID NOW COVID-19 PRODUCT INSERT INCLUDES A LIMITATION THAT FALSE NEGATIVE RESULTS MAY OCCUR IF A SPECIMEN IS IMPROPERLY COLLECTED, TRANSPORTED OR HANDLED. FALSE NEGATIVE RESULTS MAY ALSO OCCUR IF AMPLIFICATION INHIBITORS ARE PRESENT IN THE SPECIMEN OR IF INADEQUATE LEVELS OF VIRUSES ARE PRESENT IN THE SPECIMEN. DUE TO THE RISK OF A POTENTIAL FALSE NEGATIVE RESULT LEADING TO NO OR DELAYED TREATMENT, THIS EVENT SHALL BE CONSIDERED REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
360138 IDNOW COVID-19 MOLECULAR IVD FOR IDNOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M133949

Patients

Seq Age Sex Outcome Treatment
1