FDA Adverse Event Malfunction Summary report: N

CS&T IVD BEADS 50TEST

MDR report key: 11457243 · Received March 10, 2021

Report

Report Number
2916837-2021-00078
Event Type
Malfunction
Date Received
March 10, 2021
Date of Event
February 9, 2021
Report Date
March 16, 2021
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
OYE
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION OF THE COMPLAINT, IT HAS BEEN DETERMINED THIS IS NOT A REPORTABLE EVENT. INCORRECT RESULTS ON INSTRUMENT SET UP IS NOT CONSIDERED A REPORTABLE EVENT.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE CS&T IVD BEADS 50 TEST THERE WAS ERRONEOUS RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "CS&T IVD BEADS 50 TEST FALSE POSITIVE. STRONG CD7-APC SIGNALS ACTUALLY "SPILLS?/AUTOFLUORESCENCE?" INTO THE CYCD3-V450 CHANNELS AND MAKE IT AS A STRONG POSITIVE AS WELL. BUT THE T-ALL TUBE 1 RESULT ACTUALLY SHOWS THAT MAJORITY OF THE POI IS HAVING CYCD3-.".

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE CS&T IVD BEADS 50 TEST THERE WAS ERRONEOUS RESULTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "CS&T IVD BEADS 50 TEST FALSE POSITIVE. STRONG CD7-APC SIGNALS ACTUALLY "SPILLS?/AUTOFLUORESCENCE?" INTO THE CYCD3-V450 CHANNELS AND MAKE IT AS A STRONG POSITIVE AS WELL. BUT THE T-ALL TUBE 1 RESULT ACTUALLY SHOWS THAT MAJORITY OF THE POI IS HAVING CYCD3-."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
354554 CS&T IVD BEADS 50TEST NA OYE BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES 0065086

Patients

Seq Age Sex Outcome Treatment
1