FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-1997-00922
- Event Type
- Malfunction
- Date Received
- August 19, 1997
- Date of Event
- July 1, 1997
- Report Date
- July 21, 1997
- Manufacturer
- BAXTER HEALTHCARE PTE LTD.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
EVALUATION INFO: THE INFUSION PUMP WAS EVALUATED BY BAXTER HEALTHCARE. THE DEVICE UTILIZES AN ULTRASONIC SENSOR, WHERE AIR WOULD TRANSMIT THE SIGNAL POORLY AND STIMULATE AN AIR ALARM. WHEN AN AIR FILLED INTRA VENOUS TUBING WAS LOADED INTO THE PUMP THE TRANSMITTED ULTRASONIC SIGNAL DID NOT DROP BELOW THE THRESHOLD TO INITIATE AN AIR ALARM. CO'S RECORDS INDICATE THAT THE AIR SENSOR WAS SUCCESSFULLY TESTED IN MAY OF 1995. HOWEVER, FURTHER INSPECTION SHOWED NO DIRECT CAUSE FOR THE CHANGE IN THE AIR SENSOR VALUES THAT WERE OBSERVED. THE AIR SENSING SYSTEM FOR THE PUMP WAS RECALIBRATED BEFORE THE PUMP WAS RETURNED TO THE HOSP.
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE TESTING BY THE HOSPITAL BIOMEDICAL ENGINEERING DEPT, CHANNEL ONE OF THE PUMP DID NOT DETECT A 110 MICROLITER SIZE AIR GAP. THERE WAS NO CLINICAL INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE LTD. | 6301 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |