SYRINGE 5ML SALINE FILL CHINA SP
Report
- Report Number
- 1911916-2021-00190
- Event Type
- Malfunction
- Date Received
- March 9, 2021
- Date of Event
- February 8, 2021
- Report Date
- March 3, 2021
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- NGT
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 306594 AND LOT NUMBER 0063223. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT THIS TYPE OF DEFECT DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. INVESTIGATION CONCLUSION: BASED ON THE INVESTIGATION AND WITH NO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. WE WILL CONTINUE MONITORING THE COMPLAINT TREND FOR THIS PRODUCT AND SYMPTOM. WITH NO SAMPLE ANALYSIS A PROBABLE ROOT CAUSE COULD NOT BE OFFERED.
IT WAS REPORTED THAT THE SYRINGE 5ML SALINE FILL CHINA SP EXPERIENCED A BROKEN/DAMAGED PLUNGE ROD. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: AFTER THE PATIENT'S INFUSION, THE NURSE USED A FLUSH TO FLUSH THE TUBE. AFTER OPENING IT, IT WAS FOUND THAT THE PLUNGER OF THE FLUSH WAS BROKEN AND COULD NOT BE USED, SO IT WAS REPLACED IMMEDIATELY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339463 | SYRINGE 5ML SALINE FILL CHINA SP | PREFILLED SALINE SYRINGE | NGT | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 0063223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |