FDA Adverse Event Malfunction Summary report: N

DRIVER SHAFT, T4, NON-RETAINING, MINI AO

MDR report key: 11437296 · Received March 8, 2021

Report

Report Number
1220246-2021-02697
Event Type
Malfunction
Date Received
March 8, 2021
Date of Event
February 24, 2021
Report Date
June 2, 2021
Manufacturer
ARTHREX, INC.
Product Code
LXH
UDI-DI
00888867303553
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

COMPLAINT CONFIRMED. ONE UNPACKAGED AR-18700-29, BATCH 1391940 DRIVER SHAFT WAS RECEIVED FOR INVESTIGATION. VISUAL INSPECTION THROUGH THE USE OF MICRO-VU ID: 654 IDENTIFIED THAT THE DRIVE GEOMETRY AT THE DISTAL TIP WAS BOTH TWISTED AND BROKEN. MICRO-VU MEASUREMENTS REVEALED THAT APPROXIMATELY 0.09620" OF THE TIP REMAINED, WITH DESIGN TOLERANCES OUTLINING THE TIP LENGTH AS MEASURING BETWEEN 0.125"±.001". THE WHORL PATTERN VISUALIZED UNDER MAGNIFICATION ACROSS THE SURFACE OF THE DRIVE GEOMETRY INDICATED THAT THE TIP BROKE AS THE DEVICE WAS BEING TORQUED. MATERIAL ANALYSIS TESTING IDENTIFIED THAT THE RETURNED AR-18700-29 DRIVER SHAFT MET COMPOSITION SPECIFICATIONS. BASED ON THE OBSERVED CONDITION OF THE DISTAL TIP AND ALL FINDINGS NOTED, THIS EVENT IS MOST LIKELY THE RESULT OF OVER-TORQUING THE DRIVER DURING USE.

Additional Manufacturer Narrative · 1

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED, THAT DURING A PROXIMAL PHALANX FRACTURE OF THE 2ND FINGER REPAIR PROCEDURE, AS THE SURGEON WAS INSERTING THE SCREW WHEN THE TIP OF THE AR-18700-09, DRIVE SHAFT BROKE. THE BROKEN PIECE WAS REMOVED. THE SURGEON USED AN AR-18700-29, DRIVE SHAFT, FOR INSERTION OF THE LAST SCREW, THE SURGEON COULD FEEL THE TIP OF THE DRIVER WAS TWISTING AND BACKED OFF REMOVING THE AR-18700-29. THE CASE WAS COMPLETED LEAVING THE SCREW NOT FULLY SEATED.