FDA Adverse Event Malfunction Summary report: N

IMP TM 4.1MM MTX FULL, 11.5MM

MDR report key: 11434609 · Received March 8, 2021

Report

Report Number
0002023141-2021-00557
Event Type
Malfunction
Date Received
March 8, 2021
Date of Event
February 8, 2021
Report Date
June 14, 2021
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024018945
PMA / PMN Number
K132258
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

AN IMP TM 4.1MM MTX FULL, 11.5MM (TMT4B11) WAS RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED SLIGHTLY DRIED BLOOD AND A FRACTURE AT THE THREADS ABOVE THE TM JUNCTION. NO PRE-EXISTING CONDITIONS WERE NOTED ON THE PER. PICTURES OR X-RAY IMAGES WERE NOT PROVIDED. APPROPRIATE DOCUMENTATION WAS REVIEWED. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (1217563). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1217563) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PATIENT WEIGHT UNKNOWN / NOT PROVIDED. PMA/510K: K113753, K112160.

Description of Event or Problem · 1

IT WAS REPORTED THAT IMPLANT WAS LOOSE AND WHEN REMOVED, THE IMPLANT WAS SPLIT OFF WHERE TRABECULAR METAL STARTS. THE OTHER HALF IS STILL IN THE PATIENT FULLY INTEGRATED AT SITE #3. AS A RESULT OF THE EVENT NO INJURY TO THE PATIENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331662 IMP TM 4.1MM MTX FULL, 11.5MM DENTAL IMPLANT DZE ZIMMER DENTAL TMT4B11 1217563 00889024018945

Patients

Seq Age Sex Outcome Treatment
1 70 YR Required Intervention