FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS FLEX

MDR report key: 114278 · Received August 19, 1997

Report

Report Number
1527736-1997-02028
Event Type
Malfunction
Date Received
August 19, 1997
Report Date
May 23, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
KOG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

TRACKING #.44492. EES#.974907/C. H6; DAMAGED CARTRIDGE. ENDOPATH ETS FLEX: BASED UPON THE INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THE THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANISM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. THE RETURNED CARTRIDGE HAD BROKEN TOWERS, ROLLED DRIVERS, AND A DETACHED PAN. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. THE RETURNED CARTRIDGE HAD A BENT LOCKOUT TAB ON THE PAN WHICH INDICATES THAT THE INSTRUMENT'S FIRING CYCLE WAS INTERRUPTED, RELEASED, THEN RESTARTED. WHEN THIS OCCURRED, THE LOCKOUT TAB ON THE CARTRIDGE BECAME DAMAGED.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS USED DURING AN UNK PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE DEVICE WOULD NOT FIRE. THERE WAS NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS FLEX LINEAR CUTTER KOG ETHICON ENDO-SURGERY, INC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other