VITAL SIGNS® PRESSURE INFUSOR
Report
- Report Number
- 3004050971-2021-00015
- Event Type
- Injury
- Date Received
- March 7, 2021
- Date of Event
- February 8, 2021
- Report Date
- February 8, 2021
- Product Code
- KZD
- UDI-DI
- 10817357021063
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE HISTORY RECORD REVIEW SHOW ALL UNITS WERE PROCEEDED LEAKAGE TEST ACCORDING TO WORK INSTRUCTION AND THE INSPECTION RECORD SHOWED THE SAMPLING INSPECTION OF MANUFACTURED LOT PASS ALL THE INSPECTION REQUIREMENT ON VISUAL AND PACKAGING REQUIREMENT. THE MWI-51004353-AS-004/A SHOWED THE ASSEMBLY METHOD WAS WELDING THE PVC TUBE WITH FLANGE WITH USING THE RF WELDING MACHINE . IN MWI-510043XX-AS-004 REV.:B. ,WHICH EFFECTIVE ON (B)(6) 2016, CYC WAS APPLIED FOR BONDING OF TUBE FLANGE AND 8 INCH PVC TUBE INSTEAD OF USING WELDING METHOD , DUE TO THE MATERIAL INCOMPATIBILITY CONTRIBUTES TO THE ROOT CAUSE OF THE NON WELD. DUE TO THE LACK OF MATERIAL FOR EVALUATION AND APPLICATION INFORMATION, THE ACTUAL FAILURE ANALYSIS COULD NOT BE PERFORMED. THE CAUSE OF REPORTED CASE IS CURRENTLY UNDETERMINED.
AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. THEREFORE, ROOT CAUSE HAS NOT BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER REPORTED THAT THE VITAL SIGNS® PRESSURE INFUSOR HAS A LEAKING VALVE. IT HAPPENED DURING USE, AND THE MACHINE ALARMED "O2 CALIBRATION" AND "OCCLUSION." THE PATIENT HAD TO BE SEDATED FURTHER IN ORDER TO STOP THE ALARM FROM SOUNDING. HARM OR INJURY IS UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 324362 | VITAL SIGNS® PRESSURE INFUSOR | INFUSOR, PRESSURE, FOR I.V. BAGS | KZD | VITAL SIGNS® PRESSURE INFUSOR | 10817357021063 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |