FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL AMBER

MDR report key: 11425351 · Received March 5, 2021

Report

Report Number
3003152976-2021-00125
Event Type
Malfunction
Date Received
March 5, 2021
Date of Event
February 8, 2021
Report Date
April 26, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO PHOTOS OR PHYSICAL SAMPLES OF LOT 1912282 WHICH DISPLAY THE REPORTED CONDITION WERE PROVIDED FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 1912282 ,NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. DURING THE MOLDING PROCESS, POLYPROPYLENE IS MIXED WITH COLORANT TO ACHIEVE THE PROPER AMBER COLORING OF THE SYRINGE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES, NO ISSUES RELATED TO THIS INCIDENT WERE IDENTIFIED. ALTHOUGH NO DIRECT ISSUES WERE IDENTIFIED, THIS MALFUNCTION CAN OCCUR DUE TO IMPROPER COLORANT DOSING DURING THE MOLDING PROCESS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 50ML LL AMBER HAD FOREIGN MATTER INSIDE THE SYRINGE. THIS OCCURRED ON 5 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RED STREAKS IN THE PLASTIC OF THE SYRINGES.

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 50ML LL AMBER HAD FOREIGN MATTER INSIDE THE SYRINGE. THIS OCCURRED ON 5 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RED STREAKS IN THE PLASTIC OF THE SYRINGES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
316215 SYRINGE 50ML LL AMBER SYRINGE FMF BECTON DICKINSON, S.A. 1912282

Patients

Seq Age Sex Outcome Treatment
1