FDA Adverse Event Malfunction Summary report: N

SILICONE CHANNEL DRAIN WITH 1/8 IN. TROCAR, 10 FR., ROUND, 3/4 FLUTED

MDR report key: 11423610 · Received March 5, 2021

Report

Report Number
1018233-2021-01022
Event Type
Malfunction
Date Received
March 5, 2021
Date of Event
February 8, 2021
Report Date
July 21, 2021
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
GBX
UDI-DI
00801741049835
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS CONFIRMED CAUSE UNKNOWN. 1 SAMPLE WAS CONFIRMED TO EXHIBIT THE REPORTED FAILURE. THE DEVICE HAD NOT MET SPECIFICATIONS. THE PRODUCT WAS USED FOR TREATMENT. THE PRODUCT CAUSED THE REPORTED FAILURE. VISUAL EVALUATION OF THE RETURNED SAMPLE NOTED ONE OPENED (WITHIN ORIGINAL PACKAGING), UNUSED WOUND DRAIN. VISUAL INSPECTION OF THE SAMPLE NOTED THAT WOUND DRAIN TUBING WAS KNOTTED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT SHOW ANY PROBLEMS OR CONDITIONS THAT WOULD HAVE CONTRIBUTED TO THE REPORTED ISSUE. THEREFORE, NO ADDITIONAL ACTION IS REQUIRED AT THIS TIME. THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "INDICATIONS: WOUND DRAINS ARE USED TO REMOVE EXUDATES FROM WOUND SITES. WARNINGS: AN EFFECTIVE CLOSED SUCTION DRAIN SYSTEM REQUIRES MAINTENANCE OF THE SYSTEM TO PRESERVE PATENCY. THE DRAIN MUST NOT BE ALLOWED TO OCCLUDE NOR THE RESERVOIR TO COMPLETELY FILL; AND RESERVOIR SUCTION MUST BE MAINTAINED IN ORDER FOR THE SYSTEM TO FUNCTION PROPERLY. VERIFY THAT THE SYSTEM IS FUNCTIONING PROPERLY. IF THE SYSTEM IS NOT MAINTAINED PROPERLY, SURGICAL COMPLICATIONS, INCLUDING HEMATOMAS, MAY RESULT. IN THE EVENT OF OCCLUSION OF THE DRAIN, ALL WOUND DRAINAGE VIA THE DRAIN CEASES. CAREFUL ATTENTION TO THE DRAIN WILL MINIMIZE THE POSSIBILITY OF THIS PROBLEM. IF OCCLUSION DOES OCCUR, THE DRAIN CAN BE ASPIRATED BY CONNECTING SUCTION TO THE RESERVOIR OUTLET OR TEMPORARILY DISCONNECTING THE DRAIN FROM THE RESERVOIR AND APPLYING SUCTION DIRECTLY TO THE DRAIN. IF AN AIR-TIGHT SEAL BETWEEN THE DRAIN AND THE SKIN WHERE THE DRAIN EMERGES IS NOT ACHIEVED, THE AIR LEAK MUST BE RECTIFIED OR THE SYSTEM MUST BE CONVERTED TO OPEN DRAINAGE. AN AIRTIGHT SEAL BETWEEN ALL SYSTEM COMPONENTS (DRAIN, ADAPTOR AND RESERVOIR) IS NECESSARY FOR PROPER SYSTEM FUNCTION. LEAVING THE SOFT SILICONE ELASTOMER DRAIN IMPLANTED FOR ANY PERIOD OF TIME SO AS TO CAUSE TISSUE INGROWTH AROUND THE DRAIN CAN INTERFERE WITH EASY REMOVAL AND MAY EFFECT THE PERFORMANCE OF THE DRAIN. THE SURGEON SHOULD MONITOR THE PATIENT¿S RATE OF WOUND HEALING. EVACUATORS SHOULD BE USED IN CARDIO-THORACIC SURGERY ONLY AFTER THE LUNG IS FULLY EXPANDED AND ALL AIR LEAKS HAVE SEALED. DRAIN PERFORATIONS OR CHANNELS MUST LIE WITHIN THE WOUND OR CAVITY TO BE DRAINED, OTHERWISE INADEQUATE DRAINAGE MAY RESULT. TO AVOID THE POSSIBILITY OF DRAIN DAMAGE OR BREAKAGE: · AVOID SUTURING THROUGH DRAINS. · DRAINS SHOULD LIE FLAT AND IN LINE WITH THE SKIN EXIT AREAS. · PARTICULAR CARE SHOULD BE TAKEN TO AVOID ANY OBSTACLES TO THE DRAIN EXIT PATH. · DRAINS SHOULD BE CHECKED FOR FREE MOTION DURING CLOSURE TO MINIMIZE THE POSSIBILITY OF BREAKAGE. · DRAIN REMOVAL SHOULD BE DONE GENTLY BY HAND. DRAINS SHOULD NOT BE HANDLED WITH POINTED, TOOTHED OR SHARP INSTRUMENTS WHICH COULD CAUSE CUTS OR NICKS AND LEAD TO SUBSEQUENT STRUCTURAL FAILURE OF THE DRAIN. · SURGICAL REMOVAL MAY BE NECESSARY IF DRAIN IS DIFFICULT TO REMOVE OR BREAKS. COMPLICATIONS COMPLICATIONS WHICH MAY RESULT FROM THE USE OF THIS SUCTION DRAINAGE SYSTEM INCLUDE THE RISKS ASSOCIATED WITH METHODS UTILIZED IN THE SURGICAL PROCEDURE, AS WELL AS THE PATIENTS DEGREE OF INTOLERANCE TO ANY FOREIGN OBJECT IN THE BODY." CORRECTION: D,H THE ACTUAL/SUSPECTED DEVICE WAS INSPECTED

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BARD SILICONE CHANNEL DRAIN WAS KNOTTED.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BARD SILICONE CHANNEL DRAIN WAS KNOTTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
316104 SILICONE CHANNEL DRAIN WITH 1/8 IN. TROCAR, 10 FR., ROUND, 3/4 FLUTED BARD SILICONE CHANNEL DRAIN GBX C.R. BARD, INC. (COVINGTON) -1018233 072221 NGER2784 00801741049835

Patients

Seq Age Sex Outcome Treatment
1 Other