FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED STD STEM SIZE 1

MDR report key: 11420961 · Received March 5, 2021

Report

Report Number
3005180920-2021-00170
Event Type
Injury
Date Received
March 5, 2021
Date of Event
February 4, 2021
Report Date
March 5, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804076
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 16 FEBRUARY 2021: LOT 111767: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25-JULY-2011. EXPIRATION DATE: 2016-06-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT SINCE 1-1-2017.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM. THE CAUSE OF THE LOOSE STEM IS UNKNOWN. 9 YEARS AND 6 MONTHS AFTER PRIMARY THE SURGEON REVISED THE STEM, HEAD, LINER, AND CUP WITH COMPETITOR IMPLANTS. THE SURGEON DECIDED TO REVISE THE CUP BECAUSE IN THE X-RAYS, IT APPEARED LOOSE. HOWEVER, DURING SURGERY, IT WAS OBSERVED THAT NOTHING WAS WRONG WITH THE CUP, BUT THE SURGEON DECIDED TO REVISE IT ANYWAY. X-RAYS ARE NOT AVAILABLE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
315943 STEM: AMISTEM H HA COATED STD STEM SIZE 1 CEMENTLESS STEM LZO MEDACTA INTERNATIONAL SA 01.18.131 111767 07630030804076

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention