NXSTAGE SYSTEM ONE
Report
- Report Number
- 3003464075-2008-00401
- Event Type
- Injury
- Date Received
- August 28, 2008
- Date of Event
- August 2, 2008
- Report Date
- August 2, 2008
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK OF THE PT'S BLOOD DUE TO CLOTTING. EVAL OF THE RETURNED CARTRIDGE CONFIRMED A DIALYSATE LEAK AT THE BALANCE CHAMBER OF THE CARTRIDGE. THE EXACT CAUSE OF THE LEAK IS UNDER INVESTIGATION. THE USER'S GUIDE INCLUDES THE FOLLOWING WARNING, "THE NXSTAGE CYCLER MAY NOT SENSE SLOW FLUID OR BLOOD LEAKS RESULTING FROM LOOSE CONNECTIONS, FAULTY COMPONENTS, VENOUS ACCESS DISCONNECTION OR OTHER POTENTIAL CAUSES. LEAKING FLUIDS COULD LEAD TO BLOOD LOSS, INJURY, OR DEATH. LEAKING FLUIDS COULD ALSO CAUSE A PERSON TO SLIP OR FALL. ALWAYS TIGHTEN AND RECHECK PT VASCULAR ACCESS AND ALL FLUID LINES, CONNECTIONS AND CLAMPS. SECURE THE BLOOD ACCESS DEVICE TO THE PT. VISUALLY INSPECT THE SYSTEM PERIODICALLY FOR EVIDENCE OF LEAKS. TERMINATE TREATMENT IF LEAK CAN NOT BE RESOLVED." A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADD'L INFO WILL BE PROVIDED.
THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD POLICY. AN ALARM OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT INDICATING THAT THE EXTRACORPOREAL BLOOD CIRCUIT WAS CLOTTED. A CLEAR FLUID LEAK WAS OBSERVED. TREATMENT WAS DISCONTINUED WITHOUT PERFORMING RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | CAR-170 | 8037712 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Other |