NXSTAGE SYSTEM ONE

MDR report key: 1140948 · Received August 28, 2008

Report

Report Number
3003464075-2008-00404
Event Type
Injury
Date Received
August 28, 2008
Date of Event
August 3, 2008
Report Date
August 3, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK IN A TIMELY MANNER. THE USER'S GUIDE INSTRUCTS THE OPERATOR TO PROMPTLY RINSEBACK THE PT'S BLOOD IN THE EVENT OF AN UNRECOVERABLE ALARM. THE DISPOSABLE CARTRIDGE WAS NOT AVAILABLE FOR EVAL. THE EXACT CAUSE OF THE AIR ALARMS CANNOT BE DETERMINED BUT IS UNLIKELY CARTRIDGE RELATED AS THE ALARMS OCCURRED MID TREATMENT. THE USER'S GUIDE PROVIDES ADEQUATE INSTRUCTIONS FOR TROUBLESHOOTING AIR ALARMS. THERE HAVE BEEN NO SIMILAR PROBLEMS REPORTED SUBSEQUENT TO THIS EVENT. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADD'L INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. ARTERIAL AIR ALARMS OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT WHICH COULD NOT BE RESOLVED BY THE OPERATOR. PARTIAL RINSEBACK WAS COMPLETED WHEN POSSIBLE CLOTTING WAS NOTED, RESULTING IN AN ESTIMATED BLOOD LOSS OF 150CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8017709

Patients

Seq Age Sex Outcome Treatment
1 52 YR Other