NXSTAGE SYSTEM ONE

MDR report key: 1140923 · Received August 28, 2008

Report

Report Number
3003464075-2008-00397
Event Type
Injury
Date Received
August 28, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK OF THE PT'S BLOOD IN A TIMELY MANNER AS INSTRUCTED IN THE USER'S GUIDE. EVALUATION OF THE RETURNED CARTRIDGE CONFIRMED A DIALYSATE LEAK AT THE BALANCE CHAMBER OF THE CARTRIDGE. THE EXACT CAUSE OF THE LEAK IS UNDER INVESTIGATION. THE USER'S GUIDE INCLUDES THE FOLLOWING WARNING, "THE NXSTAGE CYCLER MAY NOT SENSE SLOW FLUID OR BLOOD LEAKS RESULTING FROM LOOSE CONNECTIONS, FAULTY COMPONENTS, VENOUS ACCESS DISCONNECTION OR OTHER POTENTIAL CAUSES. LEAKING FLUIDS COULD LEAD TO BLOOD LOSS, INJURY, OR DEATH. LEAKING FLUIDS COULD ALSO CAUSE A PERSON TO SLIP OR FALL. ALWAYS TIGHTEN AND RECHECK PT VASCULAR ACCESS AND ALL FLUID LINES, CONNECTIONS AND CLAMPS. SECURE THE BLOOD ACCESS DEVICE TO THE PT. VISUALLY INSPECT THE SYSTEM PERIODICALLY FOR EVIDENCE OF LEAKS. TERMINATE TREATMENT IF LEAK CAN NOT BE RESOLVED." A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A CLEAR FLUID LEAK OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT. THE RINSEBACK PROCESS WAS STARTED BUT COULD NOT BE COMPLETED DUE TO CLOTTING, RESULTING IN AN ESTIMATED BLOOD LOSS OF 200CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8037712

Patients

Seq Age Sex Outcome Treatment
1 38 YR Other