FDA Adverse Event Malfunction Summary report: N

STERRAD 200 STERILIZATION SYSTEM SINGLE-DOOR UNIT WITH STAINLESS STEEL EXTERIOR

MDR report key: 1140665 · Received August 28, 2008

Report

Report Number
2084725-2008-00539
Event Type
Malfunction
Date Received
August 28, 2008
Date of Event
August 2, 2008
Report Date
August 2, 2008
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K030429
Removal / Correction Number
Z-1630-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OTHER, OIL MIST - CAPITAL EQUIPMENT, THE UNIT WAS EVALUATED AT THE CUSTOMER SITE. THE FSE FOUND THE OIL MIST LEAK COMING FROM THE OIL MIST FILTER. THE FSE REPLACED THE OIL MIST FILTER ELEMENT AND THE SYSTEM MET SPECIFICATIONS.

Description of Event or Problem · 1

THE ASP FIELD SERVICE ENGINEER (FSE) FOUND OIL MIST COMING FROM THE CUSTOMER'S UNIT. THE CUSTOMER STATED THAT THERE WERE NO PHYSICAL COMPLAINTS REPORTED. THE FSE REPAIRED THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERRAD 200 STERILIZATION SYSTEM SINGLE-DOOR UNIT WITH STAINLESS STEEL EXTERIOR NONE MLR ADVANCED STERILIZATION PRODUCTS NA NA