THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP
Report
- Report Number
- 8010047-2021-03243
- Event Type
- Malfunction
- Date Received
- March 3, 2021
- Date of Event
- February 8, 2021
- Report Date
- March 26, 2021
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- GEI
- UDI-DI
- 04953170337574
- PMA / PMN Number
- K111202
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS RETURNED AND EVALUATED. THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THIS INFORMATION. THE DEVICE HISTORY RECORD REVIEW CONFIRMED THAT DEVICE LOT HAD NO ANOMALY IN INSPECTION. THE DEVICE ACTUAL LOT# KR103604 AND LOT# KR101472 IS THE PACKAGE LOT #. A TB-0535FC THUNDERBEAT (LOT# KR103604) WAS RECEIVED FOR EVALUATION DUE TO ¿BROKE OFF¿. THE USER¿S COMPLAINT WAS CONFIRMED. THE DEVICE WAS ATTACHED TO THE USG-400/ESG-400 AND A PROBE CHECK WAS PERFORMED; THE DEVICE DID NOT PASS THE PROBE CHECK. BOTH SWITCHES WERE CHECKED AND BOTH SWITCHES ARE FUNCTIONAL. A VISUAL INSPECTION ON THE RECEIVED CONDITION WAS PERFORMED ON THE DEVICE; THERE IS SOME TISSUE BUILDUP. THE PTFE PAD (TEFLON PAD) WAS INSPECTED AND FOUND NORMAL WEAR WITH NO METAL EXPOSED. THE DISTAL END OF THE DEVICE WAS INSPECTED UNDER A MICROSCOPE AND FOUND THE PROBE TIP IS BROKEN. THE BROKEN PIECE WAS NOT RETURNED FOR EVALUATION. THE WIPER MOVEMENT IS NORMAL. THE MOVEMENT OF THE HANDLE AND JAW IS NORMAL. THE HANDLE LOAD IS NORMAL. THE ROTATION OF THE KNOB TORQUE IS NORMAL AND SMOOTH. THE EXACT CAUSE OF THE EVENT CANNOT BE EXCLUSIVELY IDENTIFIED. HOWEVER BASED ON PAST INVESTIGATIONS, THE BROKEN PROBE POSSIBLY OCCURRED DUE TO CONTACT WITH A SURGICAL INSTRUMENT OR THE NON-INSULATED AREA OF GRASPING SECTION. THE DETAILED STEP-BY-STEP SCENARIO IS BELOW. CONTACT WITH A SURGICAL INSTRUMENT: 1. DURING OUTPUT IN SEAL & CUT MODE, THE PROBE CAME IN CONTACT WITH HARD TISSUE, METAL OR A SURGICAL INSTRUMENT. CONSEQUENTLY, A SCRATCH WAS MADE ON THE PROBE. 2. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 3. THE PROBE BROKE WITH ADDED LOAD. CONTACT WITH NON-INSULATED AREA OF GRASPING SECTION: 1. THE DISTAL END OF THE TISSUE PAD WAS WORN AWAY BECAUSE ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED WHILE GRASPING NOTHING (INCLUDING THE CASE THE USER KEPT ACTIVATING AFTER CUTTING TISSUE). 2. THE TISSUE PAD WAS WORN AWAY, CAUSING THE NON-INSULATED OF THE GRASPING SECTION TO TOUCH THE PROBE. 3. ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED UNDER THIS SITUATION, THEN THE SCRATCHES INDICATING THAT THE PROBE AND GRASPING SECTION WERE IN CONTACT WITH EACH OTHER WERE MADE. 4. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 5. THE PROBE BROKE WITH ADDED LOAD. THE INSTRUCTIONS FOR USE INCLUDES THE FOLLOWING STATEMENTS: DO NOT ACTIVATE OUTPUT IN SEAL & CUT MODE WHILE THE GRASPING SECTION IS CLOSED WITHOUT CONTACTING TISSUE OR VESSEL, OR ENSURING THAT TISSUE IS TRANSECTED. OTHERWISE, A LOCAL INCREASE OF THE TEMPERATURE DUE TO A FRICTION BETWEEN THE PROBE TIP AND THE GRASPING SECTION MAY RESULT IN VARIOUS FORMS OF DAMAGE IN THE PROBE TIP AND/OR THE TISSUE PAD, SUCH AS PREMATURE WEAR, BREAKAGE, DEFORMATION, AND/OR FALLING OFF INSIDE THE BODY CAVITY AND/OR PARTIAL SEPARATING. WHEN CUTTING AND VESSEL SEALING IS PERFORMED IN SEAL & CUT MODE, APPLY LIGHT TENSION ON THE TISSUE SO THAT USERS CAN CONFIRM IT IS TRANSECTED. ALSO, STOP ACTIVATION IMMEDIATELY AFTER TISSUE IS TRANSECTED. OTHERWISE, THE GRASPING SECTION, THE TISSUE PAD, OR THE PROBE TIP MAY BREAK AND FALL OFF, AND PARTIAL SEPARATING OF THE TISSUE PAD MAY OCCUR DUE TO A LOCAL INCREASE OF TEMPERATURE CAUSED BY THE FRICTION BETWEEN TISSUE PAD AND THE PROBE TIP DURING ACTIVATION. THE THUNDERBEAT INSTRUMENT SHOULD BE USED FOR SOFT TISSUE. DO NOT ACTIVATE OUTPUT WHILE GRASPING HARD TISSUE SUCH AS BONE OR HIGHLY CALCIFIED TISSUE, OR HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR, FORCEPS, AND OTHERS). OTHERWISE, IT MAY CAUSE THE PROBE TIP TO BE SCRATCHED OR COME INTO DIRECT CONTACT WITH THE METAL AREA OF THE GRASPING SECTION AS THE HEAT GENERATED BY THE FRICTION BETWEEN THE HARD OBJECT AND THE PROBE TIP COULD CAUSE WEAR/DEFORMING/SPLITTING/PROTRUDING/PARTIAL SEPARATING OF THE TISSUE PAD. IN TURN, THE PROBE MAY BREAK BEFORE DISPLAYING AN ERROR WINDOW OR GENERATING AN ALARM TONE. DO NOT GRASP OR LET THE PROBE TIP CONTACT HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR). ALSO, BE CAREFUL TO AVOID CONTACTING THE PROBE TIP WITH THOSE ACCIDENTALLY. PARTICULARLY DURING ACTIVATION, A SCRATCH ON THE PROBE TIP COULD OCCUR DUE TO ULTRASONIC VIBRATION, WHICH LEADS THE PROBE TIP TO BREAK AND FALL OFF INTO THE BODY CAVITY. IN ADDITION, THE HIGH-FREQUENCY (RF BIPOLAR) CURRENT FLOWS THROUGH THE METAL AND GENERATES SPARK DISCHARGE, WHICH MAY CAUSE BURNS AND DECREASE FUNCTIONALITIES.
THERE IS NO ADDITIONAL INFORMATION AVAILABLE FOR THIS EVENT AS YET. THE DEVICE IS NOT RETURNED, AS SUCH A DEFINITIVE ROOT CAUSE OF THE REPORTED COMPLAINT CANNOT BE DETERMINED AT THIS TIME. SUPPLEMENTAL REPORT(S) WILL BE FILED AS ANY INFORMATION BECOMES AVAILABLE.
AS REPORTED FOR THIS EVENT, DURING AN HYSTERECTOMY PROCEDURE THE PROBE OF THE DEVICE BROKE OFF INTO THE PATIENT. THE SURGEON RETRIEVED THE BROKEN PIECE. THERE WAS NO HARM OR ADVERSE IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 300415 | THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP | ULTRASONIC SURGICAL DEVICE | GEI | OLYMPUS MEDICAL SYSTEMS CORP. | TB-0535FC | KR103604 | 04953170337574 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |