ELEVATE SPINAL SYSTEM
Report
- Report Number
- 1030489-2021-00287
- Event Type
- Malfunction
- Date Received
- March 3, 2021
- Date of Event
- February 10, 2021
- Report Date
- June 24, 2021
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- MAX
- UDI-DI
- 00643169430716
- PMA / PMN Number
- K172199
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
PART ID# 7771128; LOT# 0790830W VISUAL AND OPTICAL INSPECTION REVEALED THE PEEK PORTION HAS BROKEN AWAY FROM THE TITANIUM PORTION. THERE ARE WITNESS MARKS ON THE BACKSIDE OF THE SPACER INDICATING THAT THE SPACER WAS IMPACTED. BOTH OF THE EARS ON THE PEEK PORTION OF THE SPACER HAVE BROKEN OFF INDICATING THAT THE IMPLANT CAME IN CONTACT WITH BONE AND WAS PUSHED BACK, CAUSING THE EARS TO BREAK OFF AND THAT SECTION OF THE IMPLANT TO SEPARATE. THIS TYPE OF DAMAGE IS CONSISTENT WITH EXCESSIVE FORCE APPLIED THE THREADING SCREW WHEN EXPANDING THE IMPLANT. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH UNKNOWN INDICATION FOR SPINAL THERAPY. EVENT OCCURRED INTRA-OP. IT WAS REPORTED THAT 11MM STANDARD ELEVATE CAGE PEEK PORTION CRACKED UPON IMPLANTATION. TI AND PEEK PIECES WERE BOTH EXPLANTED AND REPLACED WITH A 10MM DEVICE WITH NO ISSUES. DEVICE WAS IMPLANTED AND EXPLANTED ON SAME DAY ((B)(6)) 2021). NO DEVICE FRAGMENT LEFT IN PATIENT. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS REPORTED. DEVICE IS BEING RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 301467 | ELEVATE SPINAL SYSTEM | INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR | MAX | WARSAW ORTHOPEDICS | 7771128 | 0790830W | 00643169430716 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR |