FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP

MDR report key: 11402857 · Received March 3, 2021

Report

Report Number
1920898-2021-00250
Event Type
Malfunction
Date Received
March 3, 2021
Date of Event
February 4, 2021
Report Date
February 18, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00382903201228
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR NEEDLE BENT ON LOT # 0062051. INVESTIGATION SUMMARY: CUSTOMER RETURNED (5) 3/10CC, 12.7MM, 29G SYRINGES IN AN OPEN POLY BAG FROM LOT # 0062051. CUSTOMER STATES THAT THE NEEDLE WAS BENT. ALL RETURNED SYRINGES WERE EXAMINED AND ONE EXHIBITED A SLIGHTLY BENT CANNULA. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 0062051 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE SAMPLES AND/OR PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: CONFIRMED: BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE IS USER RELATED. THE CANNULA BENT DURING USE OF THE PRODUCT BY THE CUSTOMER. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CAPA/SA IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT 5 SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP HAD SCALE MARKING ISSUES DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "THE CUSTOMER REPORTED ABOUT FIVE SYRINGES FOR WHICH THE SCALE WAS MISALIGNED. (B)(4) 2021/2/18. A CUSTOMER'S MEMO SAID THAT 2 ISSUES HAVE OCCURRED. THE SCALE WAS MISALIGNED. THE NEEDLE WAS BENT. BDJ ADDED ROW#2 FOR " NEEDLE BENT".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303054 SYRINGE 0.3ML 29GA 1/2IN 7BAG 420CAS JP PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 320122 0062051 00382903201228

Patients

Seq Age Sex Outcome Treatment
1