FDA Adverse Event Injury Summary report: N

PUMP MMT-522NAS PRDGM INS V2.2 SK EN

MDR report key: 1140258 · Received August 29, 2008

Report

Report Number
3004209178-2008-00677
Event Type
Injury
Date Received
August 29, 2008
Date of Event
August 14, 2008
Report Date
August 18, 2008
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE WAS HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS. THE REPORTED BLOOD GLUCOSE READING WAS 222 MG/DL. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP WAS PROGRAMMED CORRECTLY. THE INSULIN PASSED THE PRIME TEST. THE INSULIN PUMP FAILED THE HIGH PRESSURE TEST ON THE FIRST ATTEMPT; HOWEVER, THE HIGH PRESSURE TEST WAS PERFORMED AGAIN WITH A NEW RESERVOIR AND INFUSION SET, AND THE INSULIN PUMP PASSED THE TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522NAS PRDGM INS V2.2 SK EN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization