FDA Adverse Event Malfunction Summary report: N

ALARIS SYSTEM PC UNIT

MDR report key: 11398956 · Received March 2, 2021

Report

Report Number
2016493-2021-29280
Event Type
Malfunction
Date Received
March 2, 2021
Report Date
February 22, 2021
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Removal / Correction Number
Z-2909-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS BEEN RECEIVED AND AN EVALUATION IS PENDING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETED.

Additional Manufacturer Narrative · 0

THIS DEVICE WAS EVALUATED AND REPAIRED THROUGH THE SERVICE REPAIR PROCESS. UPON VISUAL INSPECTION THE SERVICE TECHNICIAN NOTED THAT THE CORRODED U30, U31, U33, U34 ON LOGIC BOARD REPLACED LOGIC BOARD. ERROR CODE 100.5010 - REFLASHED POLO. UNRESPONSIVE BUTTONS ON KEYPAD REPLACED KEYPAD. A REVIEW OF THE DEVICE HISTORY RECORD FOR (B)(6) WAS PERFORMED, WHICH SHOWED THE DEVICE HAD A MANUFACTURE DATE OF 08SEP2016 AND CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. THE REVIEW WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE OF PRODUCT RELEASE FOR DISTRIBUTION. A REVIEW OF THE COMPLAINT HISTORY RECORD WAS PERFORMED FOR THE (B)(6) WHICH DID NOT CONFIRM SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE FOR THIS CUSTOMER.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE WAS NOT POWERING ON. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE WAS NOT POWERING ON. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH FAILURE INVESTIGATION RESULTS SHOULD THE DEVICE BE RECEIVED FOR EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS NOT POWERING ON. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
295926 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1