VDW GOLD/SILVER CONTRA ANGLE
Report
- Report Number
- 9611053-2021-00045
- Event Type
- Malfunction
- Date Received
- March 1, 2021
- Report Date
- February 26, 2021
- Manufacturer
- DENTSPLY VDW GMBH
- Product Code
- EGS
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, CALIBRATION RELATED ISSUES IN ENDO MOTORS HAVE LED TO FILE SEPARATION IN THE PAST. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. MICROMOTOR GMR1878113 (WHEN CHECKING MOTOR WE FOUND RESIDUES OF LIQUIDS OR OIL) DEFECTIVE, REPLACED WITH GMR2088551. THE BATTERY WAS REPLACED AS A PREVENTIVE MEASURE. THE HOUSING HAS CRACKED IN SEVERAL PLACES (PRESUMABLY DUE TO AN IMPACT) HAS NO EFFECT ON THE FUNCTION. CONTRA-ANGLE HANDPIECE 101193 (2017), 45396 (2012) AND FOOT CONTROL 173891 CHECKED, WITHOUT ERRORS. FUNCTION TEST AND TEST MEASUREMENTS WITHOUT ERRORS.
IN THIS EVENT IT WAS REPORTED THAT A VDW.CONTRA ANGLE 6:1 FROM VDW. GOLD RECIPROC DEVICE CAUSES ERROR CODE 2 ON DEVICE. CUSTOMER HAS 3 DIFFERENT CONTRA ANGLES AND THEY ALL CAUSES ERROR CODE 2. THIS IS FOR THE 2ND OF THREE. NO PATIENT INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 293261 | VDW GOLD/SILVER CONTRA ANGLE | HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL | EGS | DENTSPLY VDW GMBH | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |