FDA Adverse Event Injury Summary report: N

EVOLUT PRO TRANSCATHETER AORTIC VALVE

MDR report key: 11394556 · Received March 1, 2021

Report

Report Number
2025587-2021-00715
Event Type
Injury
Date Received
March 1, 2021
Date of Event
December 14, 2020
Report Date
March 1, 2021
Manufacturer
MEDTRONIC HEART VALVES DIVISION
Product Code
NPT
PMA / PMN Number
P130021
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CITATION: SINGH G, LE V, WIECHMANN RJ, SCHREITER SW. SELF-EXPANDABLE TRANSCATHETER AORTIC VALVE FRAME INFOLDING: AN INCREASINGLY RECOGNIZED COMPLICATION. EJCRIM 2020;7: DOI:10.12890/2020_002100. NO UNIQUE DEVICE IDENTIFIER (SERIAL/LOT) NUMBERS WERE PROVIDED; WITHOUT THIS INFORMATION IT COULD NOT BE DETERMINED WHETHER THESE OBSERVATIONS HAVE BEEN PREVIOUSLY REPORTED. WITHOUT RETURN OF THE PRODUCT NO DEFINITIVE CONCLUSION CAN BE MADE REGARDING THE CLINICAL OBSERVATIONS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION VIA LITERATURE REGARDING A NINETY-THREE-YEAR-OLD MALE PATIENT WITH SYMPTOMATIC SEVERE AORTIC STENOSIS AND AORTIC VALVE CALCIFICATION THAT EXPENDED FROM THE AORTIC ANNULUS TO THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT) WHO UNDERWENT THE IMPLANT OF A MEDTRONIC EVOLUTPRO TRANSCATHETER AORTIC VALVE (UNIQUE DEVICE IDENTIFIER NUMBERS NOT PROVIDED). DURING THE IMPLANT THE VALVE WAS RECAPTURED TWICE. ONCE DEPLOYED, SEVERE PARAVALVULAR LEAK (PVL), INCREASED GRADIENT MEASUREMENTS (MEAN GRADIENT 33 MILLIMETERS OF MERCURY (MMHG) AND AN INFOLD WERE NOTED. SUBSEQUENTLY, A FEW DAYS AFTER THE IMPLANT A BALLOON AORTIC VALVULOPLASTY (BAV) WAS PERFORMED WHICH DECREASED THE GRADIENT TO 11 MMHG AND IMPROVED THE PVL TO MILD. NO ADDITIONAL ADVERSE PATIENT EFFECTS OR PRODUCT PERFORMANCE ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288679 EVOLUT PRO TRANSCATHETER AORTIC VALVE AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED NPT MEDTRONIC HEART VALVES DIVISION EVOLUTPRO-23

Patients

Seq Age Sex Outcome Treatment
1 93 YR Required Intervention