FDA Adverse Event Malfunction Summary report: N

GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET

MDR report key: 11389608 · Received February 27, 2021

Report

Report Number
3002808486-2021-00197
Event Type
Malfunction
Date Received
February 27, 2021
Report Date
April 27, 2021
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: (DATE OF BIRTH). THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

INFORMATION PROVIDED BY (B)(6). OCCUPATION: NON-HEALTHCARE PROFESSIONAL. PMA/510(K) - K172557. INVESTIGATION: FILTER OR FILTER FRAGMENT MIGRATION AND (OR) EMBOLIZATION (E.G., MOVEMENT TO THE HEART OR LUNGS) HAS BEEN REPORTED. FILTER OR FILTER FRAGMENT MOVEMENT HAS OCCURRED IN BOTH THE CRANIAL AND CAUDAL DIRECTION AND MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS SMALLER OR LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; DEPLOYMENT INTO THROMBUS; DISLODGEMENT DUE TO LARGE THROMBUS BURDENS; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: FILTER MIGRATION, TRAUMA TO ADJACENT STRUCTURES. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Description of Event or Problem · 1

NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AT THIS TIME.

Description of Event or Problem · 1

IT IS ALLEGED THAT THE PATIENT RECEIVED A COOK GUNTHER TULIP FILTER AND IS ALLEGING MIGRATION. HOSPITAL AND MEDICAL RECORDS HAVE NOT BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
286350 GUNTHER TULIP NAVALIGN JUGULAR & FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE E3046051

Patients

Seq Age Sex Outcome Treatment
1