STYLE 110 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-02174
- Event Type
- Injury
- Date Received
- February 26, 2021
- Report Date
- June 10, 2021
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED ONE EXTENDED OPENING, BLACK PARTICLES MATERIAL WAS FOUND ON THE SHELL AND FLAT CREASES. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY: ONE SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS IN THE SIDE POSTERIOR ASSESSED AS SURGICAL IMPACT OPENING.
HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 284269 | STYLE 110 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2439952 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |