ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-00593
- Event Type
- Malfunction
- Date Received
- February 25, 2021
- Date of Event
- January 5, 2021
- Report Date
- April 21, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 10811877011320
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ADDITIONAL INFORMATION/CORRECTED DATA: PLEASE DISREGARD THE INVESTIGATION RESULTS REPORTED UNDER FOLLOW-UP MDR #1 FOR THIS MFR. REPORT # AS THIS MFR. REPORT # IS NOT RELATED TO/ASSIGNED TO THIS EVENT AND WAS RELATED IN ERROR FOR THE EVENTS REPORTED UNDER MFR. REPORTS: 1221359-2021-00604 1221359-2021-00586 1221359-2021-00587 1221359-2021-00588 1221359-2021-00589 1221359-2021-00590 1221359-2021-00591 1221359-2021-00592 1221359-2021-00594 1221359-2021-00595 1221359-2021-00596 1221359-2021-00599.
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 1013538 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000/ LOT 1013538, TEST BASE PART NUMBER 190-430/ LOT 1013538. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1013538 SHOWED THAT THE COMPLAINT RATE IS (B)(4). THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLES.
THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A NASOPHARYNGEAL SAMPLE. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING ON AN UNKNOWN SWAB WITH PCR GENERATED NEGATIVE RESULTS; CT VALUES NOT PROVIDED. THE CUSTOMER STATED THE PATIENT WAS NOT SYMPTOMATIC. IT WAS REPORTED THE PATIENT WAS TAKEN TO THE GOVERNMENT HOSPITAL FOR FURTHER TESTING AFTER RECEIVING A POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT. POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2 RNA; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN THE CLEANING SECTION OF THE INSTRUMENT USER MANUAL. DUE TO THE RISK OF A FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARANTINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT(S), DELAYED EMERGENCY TREATMENT AND/OR INAPPROPRIATE TREATMENT WITH ANTIVIRAL THERAPY, THE INCIDENT SHALL BE CONSIDERED REPORTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274675 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 1013538 | 10811877011320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |