FDA Adverse Event Malfunction Summary report: N

ADULT VENTILATOR DUAL HEATED CIRCUIT KIT

MDR report key: 11381116 · Received February 25, 2021

Report

Report Number
9611451-2021-00219
Event Type
Malfunction
Date Received
February 25, 2021
Report Date
January 29, 2021
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KIT IS NOT SOLD IN THE UNITED STATES OF AMERICA (USA) BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K122432. METHOD: ONE OF THE TWO COMPLAINT 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KITS WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (F&P) IN NEW ZEALAND FOR INVESTIGATION, WHERE IT WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: VISUAL INSPECTION OF THE COMPLAINT DEVICE REVEALED NO SIGNS OF DAMAGE. THE PRESSURE TEST ALSO REVEALED THAT THE BREATHING CIRCUIT WAS WITHIN SPECIFICATION. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE 950A81 BREATHING CIRCUITS TO FAIL THE LEAK TEST, AS REPORTED BY THE CUSTOMER, AS NO FAULT WAS FOUND WITH THE RETURNED DEVICE. ALL 950A81 ADULT VENTILATOR DUAL HEATED BREATHING CIRCUITS ARE 100% LEAK TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THE SUBJECT 950A81 BREATHING CIRCUIT WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT THE TIME OF PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KIT ILLUSTRATE STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM BEFORE CONNECTING TO A PATIENT." "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." THE USER INSTRUCTIONS ALSO WARN THE USER: "SET APPROPRIATE VENTILATOR OR FLOW SOURCE ALARMS TO MONITOR THERAPY DELIVERY."

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN AUSTRALIA REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT TWO 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KITS FAILED THE LEAK TEST DURING DEVICE SETUP. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

(B)(4). THE 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KIT IS NOT SOLD IN THE UNITED STATES OF AMERICA (USA) BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K122432. THE COMPLAINT 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KIT IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE (B)(6) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED, VIA A FISHER & PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE, THAT TWO 950A81 ADULT VENTILATOR DUAL HEATED CIRCUIT KITS FAILED THE LEAK TEST DURING DEVICE SETUP. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
280132 ADULT VENTILATOR DUAL HEATED CIRCUIT KIT BZE BZE 950A81 NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1