FREESTYLE LIBRE 2
Report
- Report Number
- 2954323-2021-05159
- Event Type
- Injury
- Date Received
- February 25, 2021
- Date of Event
- May 7, 2020
- Report Date
- February 25, 2021
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- QLG
- PMA / PMN Number
- K193371
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
ABBOT DIABETES CARE PRODUCT QUALITY ENGINEERING (PQE) INVESTIGATED THE ALARM-2 (SIGNAL LOSS ALARM) COMPLAINT. THE SERIAL NUMBER OF THE FREESTYLE LIBRE 2 SENSOR ASSOCIATED WITH THIS COMPLAINT IS UNKNOWN. DURING INVESTIGATION OF THE TRACK AND TREND REVIEW RELATED TO ALARM-2 CASES IN APRIL 2020 THIS FAILURE MODE WAS IDENTIFIED AND IS A KNOWN MANUFACTURING ISSUE THAT IMPACTS LIBRE 2 SENSORS MANUFACTURED AT FLEX BUFFALO GROVE (FBG). AS THE SERIAL NUMBER OF THE SENSOR ASSOCIATED WITH THIS COMPLAINT IS UNKNOWN, PQE IS UNABLE TO DETERMINE IF THE SENSOR IS IMPACTED BY THIS MANUFACTURING ISSUE. OUTSIDE OF THE KNOWN MANUFACTURING ISSUE, THE AVAILABLE TRIPPED TREND REPORTS FOR LIBRE 2 SENSOR AND ALARM-2 HAVE BEEN REVIEWED. THE REVIEW DID NOT IDENTIFY ANY ADDITIONAL TRENDS THAT WOULD INDICATE A PRODUCT RELATED ISSUE RELATED TO THIS COMPLAINT. IF THE PRODUCT IS RETURNED, THE CASE WILL BE RE-OPENED, AND A PHYSICAL INVESTIGATION WILL BE PERFORMED. THE DEVICE MANUFACTURING DATE IS UNKNOWN. THE DATE ENTERED IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
THE CUSTOMER REPORTED A SIGNAL LOSS ISSUE WITH THE ADC FREESTYLE LIBRE 2 SENSOR, WHICH RESULTED IN THE LOW GLUCOSE ALARM FAILING TO TRIGGER. THE CUSTOMER REPORTED EXPERIENCING SEIZURE AND LOSS OF CONSCIOUSNESS, AND WAS TRANSPORTED TO A HOSPITAL. THE CUSTOMER WAS DIAGNOSED WITH HYPOGLYCEMIA AND RECEIVED UNSPECIFIED TREATMENT. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 279606 | FREESTYLE LIBRE 2 | FLASH GLUCOSE MONITORING SYSTEM | QLG | ABBOTT DIABETES CARE INC | 71992-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |