FDA Adverse Event Malfunction Summary report: N

TRIOS, TABLE BASE

MDR report key: 11377827 · Received February 25, 2021

Report

Report Number
2921578-2021-00006
Event Type
Malfunction
Date Received
February 25, 2021
Date of Event
February 22, 2021
Report Date
April 6, 2021
Manufacturer
MIZUHO ORTHOPEDIC SYSTEMS, INC.
Product Code
JEA
UDI-DI
00842430103957
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

UPON COMPLETING A PREVENTATIVE MAINTENANCE THE ONLY ISSUE FOUND WERE LOOSE BOLTS THAT MOUNT THE HEAD-END COLUMN TO THE BASE OF THE TRIOS TABLE, WHICH IS NOT RELATED TO THE TABLE TILTED ON ITS OWN. IF THE TABLE TILTED AT ALL, IT WOULD HAVE TO HAVE THE FOOT-END BRAKE PHYSICALLY RELEASE BY THE STAFF, OTHERWISE THE TABLE TILT WILL NOT OCCUR. THIS LEAD TO POTENTIAL OPERATOR ERROR, INDICATING THAT SOMEONE ON THE STAFF PHYSICAL RELEASED THE 180 DEGREE HANDLE ON THE HEAD-SECTION ASSEMBLY. WHEN YOU RELEASE THE 180 DEGREE HANDLE, BOTH AMBER LIGHTS WILL BLINK, WHICH INDICATES THAT BRAKE IS RELEASE.

Description of Event or Problem · 0

IT WAS REPORTED TABLETOP STARTED TILTING AT THE END OF THE PROCEDURE, LUCKILY THE NURSE CAUGHT THE PATIENT BEFORE HE/SHE FELL.

Description of Event or Problem · 1

IT WAS REPORTED TABLETOP STARTED TILTING AT THE END OF THE PROCEDURE, LUCKILY THE NURSE CAUGHT THE PATIENT BEFORE HE/SHE FELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276870 TRIOS, TABLE BASE SURGICAL OPERATING TABLE JEA MIZUHO ORTHOPEDIC SYSTEMS, INC. 7803 00842430103957

Patients

Seq Age Sex Outcome Treatment
1