TRIOS, TABLE BASE
Report
- Report Number
- 2921578-2021-00006
- Event Type
- Malfunction
- Date Received
- February 25, 2021
- Date of Event
- February 22, 2021
- Report Date
- April 6, 2021
- Manufacturer
- MIZUHO ORTHOPEDIC SYSTEMS, INC.
- Product Code
- JEA
- UDI-DI
- 00842430103957
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
UPON COMPLETING A PREVENTATIVE MAINTENANCE THE ONLY ISSUE FOUND WERE LOOSE BOLTS THAT MOUNT THE HEAD-END COLUMN TO THE BASE OF THE TRIOS TABLE, WHICH IS NOT RELATED TO THE TABLE TILTED ON ITS OWN. IF THE TABLE TILTED AT ALL, IT WOULD HAVE TO HAVE THE FOOT-END BRAKE PHYSICALLY RELEASE BY THE STAFF, OTHERWISE THE TABLE TILT WILL NOT OCCUR. THIS LEAD TO POTENTIAL OPERATOR ERROR, INDICATING THAT SOMEONE ON THE STAFF PHYSICAL RELEASED THE 180 DEGREE HANDLE ON THE HEAD-SECTION ASSEMBLY. WHEN YOU RELEASE THE 180 DEGREE HANDLE, BOTH AMBER LIGHTS WILL BLINK, WHICH INDICATES THAT BRAKE IS RELEASE.
IT WAS REPORTED TABLETOP STARTED TILTING AT THE END OF THE PROCEDURE, LUCKILY THE NURSE CAUGHT THE PATIENT BEFORE HE/SHE FELL.
IT WAS REPORTED TABLETOP STARTED TILTING AT THE END OF THE PROCEDURE, LUCKILY THE NURSE CAUGHT THE PATIENT BEFORE HE/SHE FELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 276870 | TRIOS, TABLE BASE | SURGICAL OPERATING TABLE | JEA | MIZUHO ORTHOPEDIC SYSTEMS, INC. | 7803 | 00842430103957 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |