FDA Adverse Event Malfunction Summary report: N

VIDAS BRAHMS PROCALCITONIN 60T

MDR report key: 11374776 · Received February 24, 2021

Report

Report Number
8020790-2021-00015
Event Type
Malfunction
Date Received
February 24, 2021
Report Date
April 21, 2021
Manufacturer
BIOMERIEUX SA
Product Code
PRI
PMA / PMN Number
K162827
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT WAS INITIALLY SUBMITTED FOLLOWING NOTIFICATION FROM A CUSTOMER IN BRAZIL REGARDING TWO (2) FALSELY UNDER-ESTIMATED PROCALCITONIN (PCT) RESULTS IN ASSOCIATION WITH THE VIDAS® BRAHMS PROCALCITONIN 60T TESTS (REF. (B)(4) , LOT 1008130900) WHEN TESTING TWO (2) SEPARATE SAMPLES. THE CUSTOMER¿S SAMPLES WERE NOT AVAILABLE TO BE SUBMITTED FOR THE INVESTIGATION. ACCORDING TO THE ANALYSIS OF THE QUALITY DATA, NO ANOMALIES WERE IDENTIFIED DURING THE MANUFACTURING, CONTROL AND PACKAGING PROCESSES FOR VIDAS® BRAHMS PCT REF (B)(4) BATCH 1008130900/ 211206-0. THE INVESTIGATION UNIT LABORATORY TESTED SIX (6) INTERNAL SAMPLES ON THE RETAIN KIT OF VIDAS® BRAHMS PROCALCITONIN REF (B)(4) LOT 1008130900 / 211206-0. THE RESULTS OF THE SAMPLES COMPLIED WITH THE STANDARDS, AND WERE CLOSE TO THE RESULTS OBTAINED DURING THE RELEASE OF THE LOT. THERE WAS NO DRIFT OF THE LOT IDENTIFIED. THE CUSTOMER¿S INITIAL RESULTS OF <0.05 NG/ML INDICATED THE SAMPLE CONES (SPR - SOLID PHASE RECEPTACLE) MAY HAVE BEEN MISSING FROM BOTH INITIAL TESTS. LOCAL CUSTOMER SERVICE REQUIRED THE CUSTOMER TO COUNT THE SAMPLE CONES; HOWEVER, THIS INFORMATION HAS NOT BEEN PROVIDED. THE ROOT CAUSE OF THE COMPLAINT COULD NOT BE DETERMINED. BASED ON INTERNAL DATA AND INFORMATION PROVIDED BY THE CUSTOMER, A POTENTIAL ROOT CAUSE OF THE COMPLAINT IS OPERATOR ERROR LINKED TO THE ABSENCE OF THE SPR. THERE IS NO RECONSIDERATION OF THE PERFORMANCE OF PRODUCT VIDAS® BRAHMS PCT REF (B)(4) BATCH 1008130900/211206 TO ITS EXPECTATIONS.

Description of Event or Problem · 1

NOTE: REFERENCE 30450 IS NOT REGISTERED IN THE UNITED STATES. THE U.S. SIMILAR DEVICE IS PRODUCT REFERENCE 30450-01. A CUSTOMER FROM (B)(6) NOTIFIED BIOMÉRIEUX OF OBTAINING 2 FALSELY UNDER ESTIMATED PROCALCITONIN RESULTS IN ASSOCIATION WITH THE VIDAS® BRAHMS PROCALCITONIN 60T TESTS (REF. 30450, LOT 1008130900) WHEN TESTING 2 SEPARATE SAMPLES. THE CUSTOMER OBTAINED THE FOLLOWING RESULTS: TEST 1: SAMPLE (B)(6): VALUE < 0.05 NG/ML. SAMPLE (B)(6): VALUE < 0.05 NG/ML. TEST 2: SAMPLE (B)(6): VALUE = 52.46 NG/ML. SAMPLE (B)(6): VALUE = 2.25 NG/ML. THE CUSTOMER CONFIRMED THE MINIVIDAS® INSTRUMENT USED TO TEST THE SAMPLE PASSED WEEKLY QCV TESTING. THE INITIAL RESULTS OF < 0.05 NG/ML INDICATED THE SAMPLE CONES MAY HAVE BEEN MISSING FROM BOTH INITIAL TESTS. LOCAL CUSTOMER SERVICE REQUIRED THE CUSTOMER TO COUNT THE SAMPLE CONES; HOWEVER, THIS INFORMATION HAS NOT BEEN PROVIDED. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. BIOMERIEUX WILL INITIATE AN INTERNAL INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
267175 VIDAS BRAHMS PROCALCITONIN 60T VIDAS® BRAHMS PROCALCITONIN 60T PRI BIOMERIEUX SA 1008130900

Patients

Seq Age Sex Outcome Treatment
1