SYRINGE 10ML REG PR SALINE 10ML FILL
Report
- Report Number
- 9616657-2021-00025
- Event Type
- Malfunction
- Date Received
- February 24, 2021
- Date of Event
- January 26, 2021
- Report Date
- March 29, 2021
- Manufacturer
- BECTON, DICKINSON AND CO.
- Product Code
- FOZ
- UDI-DI
- 30382903065463
- PMA / PMN Number
- K003553
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 2/26/2021. H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 0223175. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS ISSUE, UNUSED SAMPLES WERE RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THE REPRESENTATIVE SAMPLES WERE INSPECTED AND NO ISSUES WITH PLUNGER MOVEMENT WERE IDENTIFIED WHEN THE PLUNGER WAS PRESSED DOWNWARDS. BASED ON THE INVESTIGATION RESULTS, A CAUSE RELATED TO THE MANUFACTURING PROCESS COULD NOT BE DETERMINED FOR THIS INCIDENT.
IT WAS REPORTED THAT 2 SYRINGE 10ML REG PR SALINE 10ML FILL HAD DIFFICULT PLUNGER MOVEMENT DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IT WAS REPORTED SALINE FLUSHES ARE NOT FLUSHING WHEN ATTACHED TO AN IV CATHETER. VERBATIM: BD TEAM, I SPOKE TO PRODUCT COMPLAINS () I AM RECEIVING MANY COMPLAINTS WITH THESE 10CC SALINE FLUSHES. ITEM 306546. THEY ARE NOT FLUSHING WHEN ATTACHED TO AN IV CATHETER. I AM COPYING MY CUSTOMER () LOTS: 0223175 , 0154213, 0188331. THIS NEEDS TO BE ADDRESSED IMMEDIATELY AS THEY HAVE OVER 30 LOCATIONS WITH THESE LOTS. PLEASE ADVISE DATED OF EVENTS FOR THE ATTACHED COMPLAINTS: MD DATE OF EVENT: (B)(6) 2021, DATE OF EVENT: (B)(6) 2021, DATE OF EVENT: (B)(6) 2021, DATE OF EVENT: (B)(6) 2021, (2 OCCURRENCES LOT 0154213 AND LOT 0188331) DATE OF EVENT: (B)(6) 2021. FROM CUSTOMER EMAIL 27-JAN THIS LOT# OF 10ML NS SYRINGES IS HAVING AN ISSUE WITH FLUSHING ONCE ATTACHED TO THE ANGIOCATH. IT CAN STILL ASPIRATE BUT IS UNABLE TO FLUSH. IT FLUSHES FINE WHEN ATTACHED TO THE EXTENSION SET ALONE."
(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT 2 SYRINGE 10ML REG PR SALINE 10ML FILL HAD DIFFICULT PLUNGER MOVEMENT DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IT WAS REPORTED SALINE FLUSHES ARE NOT FLUSHING WHEN ATTACHED TO AN IV CATHETER. VERBATIM: BD TEAM, I SPOKE TO PRODUCT COMPLAINS (). I AM RECEIVING MANY COMPLAINTS WITH THESE 10CC SALINE FLUSHES. ITEM 306546. THEY ARE NOT FLUSHING WHEN ATTACHED TO AN IV CATHETER. I AM COPYING MY CUSTOMER (). LOTS: 0223175, 0154213, 0188331, THIS NEEDS TO BE ADDRESSED IMMEDIATELY AS THEY HAVE OVER 30 LOCATIONS WITH THESE LOTS. PLEASE ADVISE. DATED OF EVENTS FOR THE ATTACHED COMPLAINTS: MD. (B)(6). FROM CUSTOMER EMAIL 27-JAN: THIS LOT# OF 10ML NS SYRINGES IS HAVING AN ISSUE WITH FLUSHING ONCE ATTACHED TO THE ANGIOCATH. IT CAN STILL ASPIRATE BUT IS UNABLE TO FLUSH. IT FLUSHES FINE WHEN ATTACHED TO THE EXTENSION SET ALONE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 268663 | SYRINGE 10ML REG PR SALINE 10ML FILL | SALINE, VASCULAR ACCESS FLUSH | FOZ | BECTON, DICKINSON AND CO. | 306546 | 0223175 | 30382903065463 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |