FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 11373299 · Received February 24, 2021

Report

Report Number
9617229-2021-02005
Event Type
Injury
Date Received
February 24, 2021
Date of Event
January 1, 2019
Report Date
July 9, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT THE CAPSULAR CONTRACTURE, RUPTURE AND DOUBLE CAPSULE WAS RECEIVED JUNE 03, 2021 WITH LOT NUMBER 3012805. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: CREASES FOLD, WEAR ABRASION AND WEIGHT TO THE SPEC. CLOUDY IN THE GEL OBSERVED AFTER AUTOCLAVE CYCLE. THE UNIT IS NOT RUPTURE. BASE ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV, RUPTURE, AND DOUBLE CAPSULE. DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE NOTED. DEVICE PHOTO EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED: RED PARTICLE MATERIAL ON THE SHELL, RED TISSUE. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE IV, RUPTURE, AND DOUBLE CAPSULE.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV, RUPTURE, AND DOUBLE CAPSULE. DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268400 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3012805

Patients

Seq Age Sex Outcome Treatment
1 49 YR Required Intervention