Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 20 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED RELATING TO UNDERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE TUBE CIT PLH 13X75 2.7 109 USI L-MG CE THERE WAS UNDER-FILL OR LOW DRAW OF A TUBE WITH BLOOD. THIS EVENT OCCURRED 2 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TRANSLATED TO ENGLISH. THE CUSTOMER STATED: "DURING USE, THE TUBES DO NOT FILL PROPERLY."

Additional Manufacturer Narrative · 1

INITIAL REPORTER EMAIL: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE TUBE CIT PLH 13X75 2.7 109 USI L-MG CE THERE WAS UNDER-FILL OR LOW DRAW OF A TUBE WITH BLOOD. THIS EVENT OCCURRED 2 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TRANSLATED TO ENGLISH. THE CUSTOMER STATED: "DURING USE, THE TUBES DO NOT FILL PROPERLY."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
266484 TUBE CIT PLH 13X75 2.7 109 USI L-MG CE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 0232884

Patients

Seq Age Sex Outcome Treatment
1