STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-01752
- Event Type
- Injury
- Date Received
- February 23, 2021
- Date of Event
- January 21, 2021
- Report Date
- June 23, 2021
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- UDI-DI
- 10888628000421
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: "THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON APR 12, 2017 WITH LOT 2884277. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: OPENINGS, CREASE FOLD, WEAR ABRASION. LEAK TEST WAS PERFORMED AND FOUND OPENING ON ANTERIOR SIDE. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED A STRIATED EDGE OPENING, CONSISTENT WITH THE USE OF A SURGICAL TOOL. THE FILL TEST INSPECTION WAS PERFORMED, AND THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: TWO STRIATED EDGE OPENINGS ON ANTERIOR AND RADIUS SIDE ASSESSED AS SURGICAL DAMAGE."
DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE WAS EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 264124 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 68HP-425 | 2884277 | 10888628000421 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Required Intervention |