FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 11368432 · Received February 23, 2021

Report

Report Number
9617229-2021-01752
Event Type
Injury
Date Received
February 23, 2021
Date of Event
January 21, 2021
Report Date
June 23, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000421
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: "THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON APR 12, 2017 WITH LOT 2884277. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: OPENINGS, CREASE FOLD, WEAR ABRASION. LEAK TEST WAS PERFORMED AND FOUND OPENING ON ANTERIOR SIDE. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED A STRIATED EDGE OPENING, CONSISTENT WITH THE USE OF A SURGICAL TOOL. THE FILL TEST INSPECTION WAS PERFORMED, AND THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: TWO STRIATED EDGE OPENINGS ON ANTERIOR AND RADIUS SIDE ASSESSED AS SURGICAL DAMAGE."

Description of Event or Problem · 0

DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE WAS EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264124 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 68HP-425 2884277 10888628000421

Patients

Seq Age Sex Outcome Treatment
1 35 YR Required Intervention