I-STAT PT/INR CARTRIDGE
Report
- Report Number
- 2245578-2021-00012
- Event Type
- Malfunction
- Date Received
- February 23, 2021
- Date of Event
- February 3, 2021
- Report Date
- June 8, 2021
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- GJS
- UDI-DI
- 10054749000170
- PMA / PMN Number
- K020355
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 03-JUN-2021. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE CARTRIDGE LOT PASSED RELEASE SPECIFICATIONS. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AF (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). PT/INR CARTRIDGE LOT T20313B IS ASSOCIATED WITH AN UNRELATED PREVIOUSLY IDENTIFIED DEFICIENCY FOR QUALITY CHECK CODE 19/ STAR OUTS, AS DOCUMENTED IN QUALITY RECORD 770729.
NA.
APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON (B)(6) 2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT PT/INR CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT PT/INR RESULTS OF 2.5 & 2.3 ON A FEMALE PATIENT WITH CHRONIC KIDNEY DISEASE STAGE 5. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. (B)(6). AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 263461 | I-STAT PT/INR CARTRIDGE | PT/INR CARTRIDGE | GJS | ABBOTT POINT OF CARE | NA | T20313B | 10054749000170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |