T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2021-20669
- Event Type
- Malfunction
- Date Received
- February 22, 2021
- Date of Event
- January 31, 2021
- Report Date
- February 22, 2021
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00853052007301
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. CUSTOMER REPORTED USING FIASP INSULIN. TANDEM CUSTOMER TECHNICAL SUPPORT INFORMED CUSTOMER THAT FIASP IS OFF LABEL PER THE USER GUIDE. CUSTOMER CHANGED INFUSION SET AND RESUMED INSULIN. CUSTOMERS BLOOD GLUCOSE WAS 123-240 MG/DL AT THE TIME OF EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 257515 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000096 | 00853052007301 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |