FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.5ML 31GA 6MM 10 BAG 500 SLA

MDR report key: 11362122 · Received February 22, 2021

Report

Report Number
1920898-2021-00198
Event Type
Malfunction
Date Received
February 22, 2021
Date of Event
January 15, 2021
Report Date
March 16, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0062085. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200880116] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 0.5ML 31GA 6MM 10 BAG 500 SLA HUB SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTED THAT IN THE LAST LOT PURCHASED, ONE OF THE SYRINGES SHOWED THE FOLLOWING QUALITY DEVIATION: WHEN REMOVING THE ORANGE PROTECTIVE CAP, THE NEEDLE (ALONG WITH THE PLASTIC PART) DETACHED ITSELF FROM THE BODY OF THE SYRINGE, GETTING STUCK IN THE ORANGE CAP.

Additional Manufacturer Narrative · 1

INITIAL REPORTER PHONE#: (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.5ML 31GA 6MM 10 BAG 500 SLA HUB SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTED THAT IN THE LAST LOT PURCHASED, ONE OF THE SYRINGES SHOWED THE FOLLOWING QUALITY DEVIATION: WHEN REMOVING THE ORANGE PROTECTIVE CAP, THE NEEDLE (ALONG WITH THE PLASTIC PART) DETACHED ITSELF FROM THE BODY OF THE SYRINGE, GETTING STUCK IN THE ORANGE CAP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257126 SYRINGE 0.5ML 31GA 6MM 10 BAG 500 SLA SYRINGE FMF BD MEDICAL - DIABETES CARE 0062085

Patients

Seq Age Sex Outcome Treatment
1