FDA Adverse Event Malfunction Summary report: N

FG, CINCHLOCK SS ANCHOR WITH INSERTER

MDR report key: 11346547 · Received February 18, 2021

Report

Report Number
0002936485-2021-00100
Event Type
Malfunction
Date Received
February 18, 2021
Date of Event
November 10, 2020
Report Date
March 1, 2021
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
MBI
UDI-DI
07613252633228
PMA / PMN Number
K131769
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ALLEGED FAILURE: THE SURGEON DEPLOYED 4X CINCHLOCK SS ANCHORS (CAT02462) INTO PATIENT. TWO OF THE ANCHORS FAILED. I BELIEVE THEY FAILED BECAUSE THE SUTURES BROKE THROUGH THE ANCHOR. NEITHER OF THE ANCHORS PULLED OUT OF THE BONE. UPDATE (B)(6) 2020: 30 MINUTE SURGICAL DELAY. THE FAILURE(S) IDENTIFIED IN THE INVESTIGATION IS CONSISTENT WITH THE COMPLAINT RECORD. THE PROBABLE ROOT CAUSES COULD BE: 1) INCOMPLETE LEVER ACTUATION, 2) EXCESSIVE FORCE APPLIED TO SUTURE, 3) USE OF EXCESSIVE FORCE, OR 4) USER NOT FAMILIAR WITH DEVICE USE. THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. NO PIECES WERE LEFT IN THE JOINT.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. NO PIECES WERE LEFT IN THE JOINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242271 FG, CINCHLOCK SS ANCHOR WITH INSERTER FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE MBI STRYKER ENDOSCOPY-SAN JOSE CAT02462 20072AE2 07613252633228

Patients

Seq Age Sex Outcome Treatment
1